  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88519",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH, PA, TX, WV.",
      "recall_number": "Z-2602-2021",
      "product_description": "DxA 5000 (DxA Automation System, DxA 5000 fit), \"For in Vitro Diagnostic Use\", Reference Numbers: B87352 with B57634, UDI: 15099590707118",
      "product_quantity": "123 systems",
      "reason_for_recall": "There is a potential that sample carriers (w or w/o samples) within automated sampling system  may transport samples to the outer lane  which does not contain a physical barrier to restrain samples.   Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).",
      "recall_initiation_date": "20210721",
      "center_classification_date": "20210930",
      "termination_date": "20241010",
      "report_date": "20211006",
      "code_info": "Affected US Serial Numbers: T23000077, T23000075, T23000076, T23000074, T23000069, T23000070, T23000066, T23000099, T23000096, T23000098, T23000100, T23000095, T23000097, T23000101, T23000092, T23000091, T23000090, T23000142, T23000045, T23000040, T23000039, T23000038, T23000152, T23000154, T23000155, T23000153, T23000106, T23000108, T23000107, T23000109, T23000128, T23000124, T23000125, T23000126, T23000115, T23000116, T23000113, T23000112, T23000117, T23000114, T23000132, T23000133, T23000131, T23000134, T23000064, T23000065, T23000063, T23000062, T23000163, T23000164, T23000162, T23000173, T23000169, T23000161, T23000138, T23000136, T23000137, T23000094, T23000093, T23000135, T23000141, T23000139, T23000140, T23000166, T23000168, T23000167, T23000082, T23000081, T23000080, T23000085, T23000084, T23000083, T23000143, T23000146, T23000145, T23000144"
    }
  ]
}