  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99062",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS,  KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI,  WY and the country of Barbados.",
      "recall_number": "Z-2594-2026",
      "product_description": "Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes.    CENTRAL LINE TRAY\t  \tDYNJ41089  \t  FISTULAGRAM PACK\t  \tDYNJ62827A  \t  FISTULOGRAM PACK\t  \tDYNJ47899A  \t  INVASIVE LINE INSERTION\t  \tDYNJ44123B  \t  NON VASCULAR PACK\t  \tDYNJ68037A  \t  VASCULAR ACCESS PACK-LF\t  \tDYNJ0101291F",
      "product_quantity": "113, 843 kits",
      "reason_for_recall": "Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA.  Products not initially reported in RES 98601.",
      "recall_initiation_date": "20260227",
      "center_classification_date": "20260625",
      "report_date": "20260701",
      "code_info": "DYNJ41089    UDI-DI 10884389891360    Lot 21BBB592    DYNJ62827A    UDI-DI 10193489306606    Lot 24BBN289    DYNJ47899A    UDI-DI 10889942574681    Lots 21GBI377   21IBL299   21LBE099   21LBU529   22JBV803    DYNJ44123B    UDI-DI 10195327381653    Lots 23FBR833   23GBN041   23HBW014   23JBO264   23JBW218   23KBI873    DYNJ68037A    UDI-DI 10195327003630    Lots 21JBQ304   21LBQ205   22ABN185   22DBT349   22EBL611   22HBH550   22IBN880   22JBH689   22KBS987   22OBA368   23ABD642   23BBH830   23DBL865   23EBT367   23FBM946   23GBT261    DYNJ0101291F    UDI-DI 10195327286644    Lots 22KDB751   23EDA119   23EDB388   23FDA396   23GDA313   23HDA759   23LDB909   24ADB667   24GDA294   24HDB515"
    }
  ]
}