      "status": "Completed",
      "city": "Danvers",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82525",
      "recalling_firm": "Medtronic Vascular",
      "address_1": "35 Cherry Hill Dr # 37a",
      "address_2": "N/A",
      "postal_code": "01923-2565",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution to US, Austria, Belgium, Bulgaria, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Italy, Kazakhstan, Luxembourg, Moldova, Netherlands, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Taiwan, and United Kingdom",
      "recall_number": "Z-2588-2019",
      "product_description": "SHERPA NX ACTIVE GUIDING CATHETER, 6F, JL3.5, 110CM, 070\", REF SA6JL35A.   for cardiovascular use",
      "product_quantity": "12 units",
      "reason_for_recall": "There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.",
      "recall_initiation_date": "20190315",
      "center_classification_date": "20190930",
      "termination_date": "20240426",
      "report_date": "20191009",
      "code_info": "GTIN: 00613994822222  All Lot/Serial Numbers"
    }
  ]
}