  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95069",
      "recalling_firm": "Inari Medical - Oak Canyon",
      "address_1": "6001 Oak Cyn",
      "address_2": "",
      "postal_code": "92618-5200",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of PA, FL, GA, CA, IL, SC, NY, MN, UT, TN, WI, MI, KS, SD, AZ, AR, IN, KY, TX, AL, MS, DE, MA, OH, MO, WV, VA, NC, NV, CO, NE, OR, NM, NH, NJ, CT, ID, IA, WA, MD, OK, RI, DC, PR, LA, ND, VT and the countries of New Zealand, Chile, Australia.",
      "recall_number": "Z-2587-2024",
      "product_description": "ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use   ClotTriever XL Catheter ClotTriever Thrombectomy System, Reference: IU-01011",
      "product_quantity": "2488",
      "reason_for_recall": "Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.",
      "recall_initiation_date": "20240719",
      "center_classification_date": "20240821",
      "report_date": "20240828",
      "code_info": "UDI-DI:00850291007277. All lots. IU-01011 REV. C.",
      "more_code_info": ""
    }
  ]
}