      "country": "United States",
      "city": "Atlanta",
      "address_1": "400 Perimeter Center Ter Ne",
      "reason_for_recall": "Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.",
      "address_2": "Ste 50",
      "product_quantity": "928 units",
      "code_info": "Versions 3.20.00 and higher",
      "center_classification_date": "20140904",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, FL, IL, IN, MD, MA, MI, MN, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, Puerto Rico; and countries of: Algeria, Angola, Antigua & Barbuda, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sir Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Ukraine.",
      "state": "GA",
      "product_description": "Monaco     The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.",
      "report_date": "20140910",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Elekta, Inc.",
      "recall_number": "Z-2581-2014",
      "initial_firm_notification": "Other",
      "product_type": "Devices",
      "event_id": "68985",
      "termination_date": "20210915",
      "more_code_info": "",
      "recall_initiation_date": "20140818",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}