  "results": [
    {
      "status": "Ongoing",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95020",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Sweden, United Kingdom.",
      "recall_number": "Z-2580-2024",
      "product_description": "VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2",
      "product_quantity": "18,467 units",
      "reason_for_recall": "It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in \"TM5-4MB\" condition codes which lead to a delay of results.",
      "recall_initiation_date": "20240703",
      "center_classification_date": "20240812",
      "report_date": "20240821",
      "code_info": "Model No 1513266; UDI/GTIN# 110758750009237; Lots within expiry: 3380, 3390, 3400, 3410, 3422, 3430, 3440, 3450, 3461, 3471, 3480.",
      "more_code_info": ""
    }
  ]
}