  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Saint Charles",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88676",
      "recalling_firm": "Elekta Inc",
      "address_1": "1450 Beale St Ste 205",
      "address_2": "N/A",
      "postal_code": "63303",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.",
      "recall_number": "Z-2563-2021",
      "product_description": "Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.",
      "product_quantity": "214 units",
      "reason_for_recall": "Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.",
      "recall_initiation_date": "20210914",
      "center_classification_date": "20210924",
      "report_date": "20211006",
      "code_info": "Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10; UDI GTIN: (01)00858164002190(10) 5.40.00, (01)00858164002190(10) 5.40.01, (01)00858164002190(10) 5.40.02, (01)00858164002268(10) 5.51.10"
    }
  ]
}