{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Marlborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99064",
      "recalling_firm": "IPG Medical Corporation",
      "address_1": "225 Cedar Hill St",
      "address_2": "N/A",
      "postal_code": "01752-5900",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of MA.",
      "recall_number": "Z-2554-2026",
      "product_description": "Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.",
      "product_quantity": "198 units",
      "reason_for_recall": "Software anomaly that causes a false display of error code 5018.",
      "recall_initiation_date": "20260505",
      "center_classification_date": "20260624",
      "report_date": "20260701",
      "code_info": "Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051, MDUF250052, MDUF250053, MDUF250054, MDUF250055, MDUF250056, MDUF250057, MDUF250058, MDUF250059, MDUF250062, MDUF250063, MDUF250064, MDUF250065, MDUF250066, MDUF250070, MDUF250072, MDUF250073, MDUF250074, MDUF250075, MDUF250084, MDUF250085, MDUF250088, MDUF250089, MDUF250090, MDUF250093, MDUF250102, MDUF250103, MDUF250105, MDUF250132, MDUF250145, MDUF250158, MDUF250161, MDUF250162, MDUF250198, MDUF250201, MDUF250206, MDUF250207, MDUF250220, MDUF250237, MDUF250238, MDUF250239, MDUF250243, MDUF250245, MDUF250279, MDUF250280, MDUF250295, MDUF250296, MDUF250298, MDUF250310, MDUF250311, MDUF250313, MDUF250323, MDUF250344, MDUF250348, MDUF250349, MDUF250350, MDUF250353, MDUF250357, MDUF250364, MDUF250365, MDUF250369, MDUF250371, MDUF250372, MDUF250378, MDUF250379, MDUF250381, MDUF250382, MDUF250383, MDUF250386, MDUF250393, MDUF250396, MDUF250397, MDUF250398, MDUF250407, MDUF250408, MDUF250411, MDUF250412, MDUF250428, MDUF250432, MDUF260016, MDUF260018, MDUF260020, MDUF260031, MDUF260041, MDUF260044, MDUF260069, MDUF260071, MDUF260078, MDUF260088, MDUF260089, MDUF260090, MDUF260091, MDUF260092, MDUF260093, MDUF250087, MDUF250208, MDUF250244, MDUF250324, MDUF250326, MDUF250354, MDUF250356, MDUF250394, MDUF260042, MDUF260045, MDUF260076; Software Version 2.2."
    }
  ]
}