  "results": [
    {
      "status": "Terminated",
      "city": "Allentown",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88531",
      "recalling_firm": "B. Braun Medical, Inc.",
      "address_1": "901 Marcon Blvd",
      "address_2": "N/A",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the country of Canada.",
      "recall_number": "Z-2545-2021",
      "product_description": "Outlook Pump Sets w/3 CARESITE Needle-Free Injection Sites: single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids  Material ID: 352413",
      "product_quantity": "585,576 units",
      "reason_for_recall": "Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.",
      "recall_initiation_date": "20210810",
      "center_classification_date": "20210924",
      "termination_date": "20240916",
      "report_date": "20211006",
      "code_info": "Lot Code: 0061755638 0061757839 0061759426 0061763839 0061763893 0061763894 0061766190 0061767014 0061767027 0061768466 0061769094 0061772192 0061772193 0061775455 0061776025 0061776298 0061776299 0061776300 0061776301 0061777154 00VL752555 00VL754131  DI: 04046964182211"
    }
  ]
}