  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Machesney Park",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80510",
      "recalling_firm": "Forte Automation Systems Inc",
      "address_1": "8155 Burden Rd",
      "address_2": "N/A",
      "postal_code": "61115-8208",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US in the state of California",
      "recall_number": "Z-2537-2018",
      "product_description": "Patient Positioning System with KRC2 controller using software versions 2.3.1   2.3.10    Product Usage:  The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.",
      "product_quantity": "13",
      "reason_for_recall": "Communications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.",
      "recall_initiation_date": "20170905",
      "center_classification_date": "20180724",
      "termination_date": "20200413",
      "report_date": "20180801",
      "code_info": "PPS Serial Number             Cabinet Type         Software Version    7757-11                               KRC2                     2.3.3     7757-21                               KRC2                     2.3.3    7757-31                               KRC2                     2.3.3     7757-41                               KRC2                     2.3.3   7757-51                               KRC2                     2.3.8    7755-21                               KRC2                     2.3.3   7755-31                               KRC2                     2.3.3   7755-41                               KRC2                     2.3.3   7392-2                                 KRC2                     2.3.3   7392-3                                 KRC2                     2.3.3    7392-4                                 KRC2                     2.3.3   7392-1                                 KRC2                     2.3.8   7392-5                                 KRC2                     2.3.8"
    }
  ]
}