  "results": [
    {
      "status": "Terminated",
      "city": "Keene",
      "state": "NH",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68739",
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "address_1": "10 Bowman Dr",
      "address_2": "N/A",
      "postal_code": "03431-5043",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to CA, CY, GB, IE, IT, JE, KW, MT, PL, RU, SA, ZA.",
      "recall_number": "Z-2537-2014",
      "product_description": "H4001S Neonatal/Paed.Intubation Kit 2.5mm Murphy,15mm Conn, Holder     Pediatric Endotracheal Intubation Kit",
      "product_quantity": "135,518 (110,246 nationwide)",
      "reason_for_recall": "The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a  suction catheter.",
      "recall_initiation_date": "20140707",
      "center_classification_date": "20140828",
      "termination_date": "20210526",
      "report_date": "20140903",
      "code_info": "Lot Numbers: 2470637, 2470636, 2457747, 2448649, 2443371, 2438610, 2438609, 2428204, 2428203, 2423483, 2412132, 2405733, 2405732, 2394049, 2394048, 2394047, 2389237, 2389236, 2385647, 2380340, 2376638, 2376473, 2371941, 2365534, 2361589, 2357817, 2357816, 2353342, 2347040, 2343290, 2339700, 2335460, 2326892, 2322812, 2309462, 2298736, 2298735, 2286584, 2278061, 2263550, 2260549, 2250041, 2233738, 2228354, 2219206, 2211922, 2208179,   2191911, 2188447, 2179221, 2175047, 2154527, 2152029, 2142174, 2128628, 2119944, 2109010, 2099495, 2090213, 2068117, 2063968, 2063967, 2042403, 12049, 12047, 12044, 11669, 11403"
    }
  ]
}