{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Reno",
      "state": "NV",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92873",
      "recalling_firm": "Hamilton Medical, Inc.",
      "address_1": "4655 Aircenter Cir",
      "address_2": "",
      "postal_code": "89502-5948",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US nationwide, including Puerto Rico.  OUS: Not provided",
      "recall_number": "Z-2529-2023",
      "product_description": "Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator",
      "product_quantity": "6 units",
      "reason_for_recall": "Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.",
      "recall_initiation_date": "20230726",
      "center_classification_date": "20230920",
      "report_date": "20230927",
      "code_info": "UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595",
      "more_code_info": ""
    }
  ]
}