  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Kuroishi",
      "state": "N/A",
      "country": "Japan",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92784",
      "recalling_firm": "Aomori Olympus Co., Ltd.",
      "address_1": "2 Chome 248-1",
      "address_2": "Okkonoki",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2522-2023",
      "product_description": "Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope.  Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012",
      "product_quantity": "1,048 units",
      "reason_for_recall": "The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002",
      "recall_initiation_date": "20230727",
      "center_classification_date": "20230907",
      "report_date": "20230913",
      "code_info": "(1)K-001: Lot Numbers: 5835330 0XI, 0YI, 12I, 14I, 15I, 16I, 17I, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK, 31K, 32K. UDI: 4953170048562 (2) K-002  Lot Numbers: 5835430 0XI, 0YI, 12I, 13I, 14I, 16I, 17I, 27K, 2XK, 2YK, 2ZK 4953170048579 K-003 5835530 0XI, 0YI, 12I, 13I, 14I, 15I, 16I, 17I, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK UDI:4953170048586 (3)K-004 Lot Numbers: 5835630 16I, 17I, 26K  UDI: 4953170048593 (4)K-011 Lot Numbers: N2485230 0XI, 15I, 17I  UDI: 4953170214035 (5) K-012  Lot Numbers: N2485330 16I, 17I  UDI: 4953170214042"
    }
  ]
}