      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management\", Atrial Lead Position Check\", EffectivCRT\" algorithms, and AdaptivCRT\").  This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.",
      "address_2": "",
      "product_quantity": "1025 devices",
      "code_info": "serial numbers:  RFA001342A  RFA001412A  RFA001748A  RFA001760A  RFA001802A  RFA001845A  RFA001868A  RFA001885A  RFA002011A  RFA002039A  RFA002152A  RFA003642A  RFA003644A  RFA003652A  RFA003658A  RFA003669A  RFA003671A  RFA003673A  RFA003677A  RFA003684A  RFA003690A  RFA003691A  RFA003692A  RFA003703A  RFA003704A  RFA003705A  RFA004075A  RFA004079A  RFA004082A  RFA004208A  RFA004218A  RFA004232A  RFA004277A  RFA004386A  RFA004390A  RFA004391A  RFA004681A  RFA004706A  RFA005006A  RFA005009A  RFA005011A  RFA005023A  RFA005025A  RFA005028A  RFA005029A  RFA005049A  RFA005079A  RFA005081A  RFA005093A  RFA005095A  RFA005097A  RFA005103A  RFA005105A  RFA005125A  RFA005249A  RFA005989A  RFA006832A  RFA008087A  RFA008119A  RFA008908A  RFA009323A  RFA010721A  RFA010722A  RFA010895A  RFA011167A  RFA011196A  RFA011348A  RFA011359A  RFA011456A  RFA011493A  RFA011519A  RFA011725A  RFA011764A  RFA011770A  RFA011774A  RFA011780A  RFA011795A  RFA011796A  RFA011797A  RFA011824A  RFA011827A  RFA011840A  RFA011846A  RFA011859A  RFA011860A  RFA011862A  RFA011872A  RFA011879A  RFA011885A  RFA011898A  RFA011919A  RFA011927A  RFA011928A  RFA011931A  RFA011933A  RFA011938A  RFA011952A  RFA011958A  RFA011959A  RFA011966A  RFA011973A  RFA011976A  RFA011978A  RFA012031A  RFA012034A  RFA012048A  RFA012124A  RFA012126A  RFA012129A  RFA012133A  RFA012149A  RFA012155A  RFA012158A  RFA012159A  RFA012160A  RFA012164A  RFA012167A  RFA012174A  RFA012176A  RFA012188A  RFA012198A  RFA012199A  RFA012200A  RFA012201A  RFA012202A  RFA012205A  RFA012223A  RFA012227A  RFA012230A  RFA012233A  RFA012234A  RFA012235A  RFA012260A  RFA012273A  RFA012274A  RFA012276A  RFA012277A  RFA012279A  RFA012290A  RFA012301A  RFA012309A  RFA012328A  RFA012369A  RFA012411A  RFA012412A  RFA012423A  RFA012433A  RFA012441A  RFA012449A  RFA012450A  RFA012469A  RFA012476A  RFA012496A  RFA012506A  RFA012521A  RFA012527A  RFA012530A  RFA012531A  RFA012532A  RFA012535A  RFA012539A  RFA012542A  RFA012554A  RFA012556A  RFA012562A  RFA012571A  RFA012580A  RFA012586A  RFA012587A  RFA012590A  RFA012598A  RFA012603A  RFA012611A  RFA012624A  RFA012633A  RFA012636A  RFA012639A  RFA012643A  RFA012647A  RFA012654A  RFA012658A  RFA012661A  RFA012667A  RFA012676A  RFA012678A  RFA012679A  RFA012693A  RFA012696A  RFA012707A  RFA012708A  RFA012719A  RFA012721A  RFA012727A  RFA012754A  RFA012757A  RFA012759A  RFA012763A  RFA012768A  RFA012771A  RFA012781A  RFA012786A  RFA012791A  RFA012802A  RFA012805A  RFA012807A  RFA012812A  RFA012813A  RFA012831A  RFA012882A  RFA012976A  RFA013056A  RFA013060A  RFA013063A  RFA013066A  RFA013067A  RFA013069A  RFA013070A  RFA013074A  RFA013086A  RFA013090A  RFA013093A  RFA013099A  RFA013105A  RFA013121A  RFA013127A  RFA013129A  RFA013162A  RFA013163A  RFA013164A  RFA013165A  RFA013167A  RFA013168A  RFA013169A  RFA013170A  RFA013171A  RFA013172A  RFA013174A  RFA013175A  RFA013201A  RFA013203A  RFA013204A  RFA013207A  RFA013208A  RFA013209A  RFA013210A  RFA013211A  RFA013213A  RFA013227A  RFA013228A  RFA013233A  RFA013235A  RFA013237A  RFA013239A  RFA013251A  RFA013255A  RFA013264A  RFA013265A  RFA013267A  RFA013269A  RFA013271A  RFA013274A  RFA013275A  RFA013277A  RFA013278A  RFA013279A  RFA013301A  RFA013303A  RFA013314A  RFA013328A  RFA013330A  RFA013388A  RFA013412A  RFA013414A  RFA013435A  RFA013453A  RFA013458A  RFA013465A  RFA013468A  RFA013490A  RFA013494A  RFA013516A  RFA013537A  RFA013557A  RFA013562A  RFA013566A  RFA013576A  RFA013584A  RFA013612A  RFA013719A  RFA013721A  RFA013730A  RFA013732A  RFA013734A  RFA013735A  RFA013745A  RFA013746A  RFA013754A  RFA013780A  RFA013787A  RFA013794A  RFA013800A  RFA013830A  RFA013832A  RFA013833A  RFA013888A  RFA013895A  RFA013901A  RFA013904A  RFA013905A  RFA013946A  RFA013952A  RFA013962A  RFA013981A  RFA014454A  RFA014479A  RFA014508A  RFA014527A  RFA014693A  RFA014710A  RFA014762A  RFA014772A  RFA014778A  RFA014781A  RFA014785A  RFA014787A  RFA014809A  RFA014810A  RFA014813A  RFA014816A  RFA014829A  RFA014837A  RFA014848A  RFA014854A  RFA015059A  RFA015060A  RFA015068A  RFA015074A  RFA015076A  RFA015099A  RFA015103A  RFA015107A  RFA015112A  RFA015133A  RFA015145A  RFA015149A  RFA015156A  RFA015160A  RFA015165A  RFA015178A  RFA015179A  RFA015186A  RFA015187A  RFA015198A  RFA015201A  RFA015202A  RFA015204A  RFA015207A  RFA015218A  RFA015223A  RFA015231A  RFA015271A  RFA015272A  RFA015276A  RFA015281A  RFA015290A  RFA015292A  RFA015297A  RFA015298A  RFA015299A  RFA015304A  RFA015311A  RFA015313A  RFA015320A  RFA015322A  RFA015326A  RFA015338A  RFA015339A  RFA015347A  RFA015348A  RFA015349A  RFA015352A  RFA015356A  RFA015357A  RFA015361A  RFA015364A  RFA015401A  RFA015402A  RFA015403A  RFA015410A  RFA015415A  RFA015417A  RFA015418A  RFA015419A  RFA015423A  RFA015427A  RFA015428A  RFA015429A  RFA015430A  RFA015438A  RFA015439A  RFA015443A  RFA015448A  RFA015455A  RFA015459A  RFA015465A  RFA015467A  RFA015468A  RFA015470A  RFA015471A  RFA015472A  RFA015480A  RFA015486A  RFA015493A  RFA015497A  RFA015498A  RFA015499A  RFA015501A  RFA015502A  RFA015567A  RFA015568A  RFA015573A  RFA015574A  RFA015576A  RFA015577A  RFA015579A  RFA015581A  RFA015582A  RFA015587A  RFA015588A  RFA015590A  RFA015592A  RFA015595A  RFA015598A  RFA015600A  RFA015601A  RFA015610A  RFA015611A  RFA015614A  RFA015616A  RFA015620A  RFA015621A  RFA015622A  RFA015624A  RFA015630A  RFA015631A  RFA015633A  RFA015635A  RFA015638A  RFA015647A  RFA015648A  RFA015651A  RFA015654A  RFA015659A  RFA015660A  RFA015661A  RFA015664A  RFA015708A  RFA015725A  RFA015726A  RFA015746A  RFA015747A  RFA015748A  RFA015751A  RFA015758A  RFA015759A  RFA015760A  RFA015761A  RFA015763A  RFA015765A  RFA015766A  RFA015771A  RFA015772A  RFA015776A  RFA015779A  RFA015783A  RFA015785A  RFA015790A  RFA015841A  RFA015842A  RFA015854A  RFA015856A  RFA015857A  RFA015873A  RFA015910A  RFA015956A  RFA015959A  RFA015962A  RFA015964A  RFA015965A  RFA015967A  RFA015968A  RFA015969A  RFA015975A  RFA015976A  RFA015981A  RFA016026A  RFA016029A  RFA016031A  RFA016032A  RFA016036A  RFA016041A  RFA016043A  RFA016056A  RFA016057A  RFA016061A  RFA016063A  RFA016065A  RFA016070A  RFA016072A  RFA016073A  RFA016079A  RFA016081A  RFA016082A  RFA016087A  RFA016103A  RFA016104A  RFA016107A  RFA016118A  RFA016123A  RFA016126A  RFA016128A  RFA016139A  RFA016142A  RFA016146A  RFA016194A  RFA016196A  RFA016197A  RFA016201A  RFA016204A  RFA016207A  RFA016210A  RFA016214A  RFA016216A  RFA016220A  RFA016227A  RFA016236A  RFA016239A  RFA016242A  RFA016244A  RFA016249A  RFA016251A  RFA016255A  RFA016256A  RFA016258A  RFA016259A  RFA016265A  RFA016272A  RFA016275A  RFA016276A  RFA016279A  RFA016288A  RFA016291A  RFA016294A  RFA016307A  RFA016312A  RFA016347A  RFA016351A  RFA016388A  RFA016389A  RFA016489A  RFA016490A  RFA016491A  RFA016494A  RFA016506A  RFA016510A  RFA016513A  RFA016514A  RFA016518A  RFA016520A  RFA016524A  RFA016528A  RFA016533A  RFA016534A  RFA016541A  RFA016544A  RFA016545A  RFA016547A  RFA016555A  RFA016556A  RFA016560A  RFA016563A  RFA016566A  RFA016567A  RFA016571A  RFA016574A  RFA016577A  RFA016586A  RFA016588A  RFA016596A  RFA016636A  RFA016641A  RFA016647A  RFA016762A  RFA016766A  RFA016767A  RFA016769A  RFA016772A  RFA016776A  RFA016777A  RFA016779A  RFA016780A  RFA016781A  RFA016783A  RFA016793A  RFA016795A  RFA016816A  RFA016825A  RFA016834A  RFA016838A  RFA016839A  RFA016848A  RFA016852A  RFA016870A  RFA016872A  RFA016878A  RFA016885A  RFA016889A  RFA016897A  RFA016905A  RFA016911A  RFA016920A  RFA016923A  RFA016926A  RFA016928A  RFA016931A  RFA016932A  RFA016933A  RFA016934A  RFA016937A  RFA016940A  RFA016942A  RFA016945A  RFA016955A  RFA016956A  RFA016958A  RFA017028A  RFA017030A  RFA017039A  RFA017040A  RFA017045A  RFA017054A  RFA017060A  RFA017063A  RFA017074A  RFA017091A  RFA017116A  RFA017119A  RFA017125A  RFA017138A  RFA017149A  RFA017166A  RFA017183A  RFA017185A  RFA017186A  RFA017196A  RFA017197A  RFA017203A  RFA017205A  RFA017206A  RFA017207A  RFA017252A  RFA017260A  RFA017266A  RFA017270A  RFA017272A  RFA017273A  RFA017277A  RFA017600A  RFA017603A  RFA017607A  RFA017611A  RFA017620A  RFA017626A  RFA017627A  RFA017632A  RFA017634A  RFA017635A  RFA017643A  RFA017647A  RFA017650A  RFA017652A  RFA017655A  RFA017658A  RFA017674A  RFA017676A  RFA017685A  RFA017686A  RFA017687A  RFA017688A  RFA017694A  RFA017698A  RFA017704A  RFA017719A  RFA017720A  RFA017721A  RFA017955A  RFA017956A  RFA017959A  RFA017964A  RFA017965A  RFA017971A  RFA017976A  RFA017980A  RFA017981A  RFA017984A  RFA017994A  RFA017996A  RFA018004A  RFA018005A  RFA018007A  RFA018010A  RFA018011A  RFA018020A  RFA018023A  RFA018024A  RFA018026A  RFA018029A  RFA018030A  RFA018033A  RFA018035A  RFA018036A  RFA018039A  RFA018042A  RFA018044A  RFA018048A  RFA018049A  RFA018054A  RFA018058A  RFA018059A  RFA018060A  RFA018062A  RFA018064A  RFA018083A  RFA018089A  RFA018091A  RFA018092A  RFA018093A  RFA018122A  RFA018128A  RFA018133A  RFA018137A  RFA018141A  RFA018144A  RFA018146A  RFA018148A  RFA018166A  RFA018173A  RFA018176A  RFA018188A  RFA018245A  RFA018252A  RFA018257A  RFA018258A  RFA018261A  RFA018262A  RFA018271A  RFA018274A  RFA018278A  RFA018283A  RFA018294A  RFA018300A  RFA018301A  RFA018305A  RFA018307A  RFA018312A  RFA018316A  RFA018319A  RFA018320A  RFA018322A  RFA018323A  RFA018326A  RFA018327A  RFA018332A  RFA018333A  RFA018340A  RFA018346A  RFA018399A  RFA018400A  RFA018401A  RFA018408A  RFA018409A  RFA018410A  RFA018416A  RFA018417A  RFA018425A  RFA018454A  RFA018459A  RFA018461A  RFA018470A  RFA018483A  RFA018486A  RFA018488A  RFA018524A  RFA018525A  RFA018526A  RFA018527A  RFA018600A  RFA018602A  RFA018604A  RFA018607A  RFA018614A  RFA018619A  RFA018623A  RFA018628A  RFA018631A  RFA018633A  RFA018654A  RFA018659A  RFA018663A  RFA018673A  RFA018805A  RFA018807A  RFA018897A  RFA018898A  RFA018899A  RFA018907A  RFA018908A  RFA018909A  RFA018913A  RFA018914A  RFA018932A  RFA018933A  RFA018934A  RFA018941A  RFA018945A  RFA018947A  RFA018950A  RFA018954A  RFA018963A  RFA018969A  RFA018970A  RFA018972A  RFA018975A  RFA018978A  RFA018979A  RFA018982A  RFA018983A  RFA018990A  RFA018992A  RFA018993A  RFA018995A  RFA018998A  RFA019005A  RFA019006A  RFA019007A  RFA019009A  RFA019012A  RFA019013A  RFA019015A  RFA019017A  RFA019020A  RFA019023A  RFA019028A  RFA019031A  RFA019032A  RFA019033A  RFA019034A  RFA019035A  RFA019042A  RFA019044A  RFA019045A  RFA019046A  RFA019047A  RFA019049A  RFA019084A  RFA019088A  RFA019189A  RFA019200A  RFA019219A  RFA019236A  RFA019247A  RFA019248A  RFA019262A  RFA019313A  RFA019361A  RFA019377A  RFA019387A  RFA019415A  RFA019417A  RFA019418A  RFA019446A  RFA019448A  RFA019453A  RFA019466A  RFA019501A  RFA019504A  RFA019651A  RFA019668A  RFA019727A  RFA019758A  RFA019957A  RFA019959A  RFA019960A  RFA019970A  RFA019971A  RFA019972A  RFA019973A  RFA019974A  RFA019975A  RFA019976A  RFA019977A  RFA019978A  RFA019982A  RFA019985A  RFA020006A  RFA020025A  RFA020027A  RFA020037A  RFA020043A  RFA020046A  RFA020051A  RFA020061A  RFA020062A  RFA020080A  RFA020104A  RFA020118A  RFA020120A  RFA020127A  RFA020139A  RFA020164A  RFA020176A  RFA020182A  RFA020188A  RFA020189A  RFA020194A  RFA020231A  RFA020234A  RFA020236A  RFA020248A  RFA020269A  RFA020277A  RFA020289A  RFA020290A  RFA020293A  RFA020300A  RFA020301A  RFA020308A  RFA020315A  RFA020325A  RFA020329A  RFA020342A  RFA020349A  RFA020356A  RFA020367A  RFA020369A  RFA020374A  RFA020383A  RFA020390A  RFA020396A  RFA020399A  RFA020401A  RFA020405A  RFA020439A  RFA020444A  RFA020448A  RFA020459A  RFA020460A  RFA020463A  RFA020469A  RFA020470A  RFA020472A  RFA020475A  RFA020481A  RFA020491A  RFA020507A  RFA020508A  RFA020513A  RFA020516A  RFA020518A  RFA020553A  RFA020555A  RFA020556A  RFA020557A  RFA020558A  RFA020735A  RFA020736A  RFA020737A  RFA020738A  RFA020739A  RFA020740A  RFA020742A  RFA020743A  RFA020775A  RFA020777A  RFA020787A  RFA020791A  RFA020792A  RFA020912A  RFA020913A  RFA020914A  RFA020915A  RFA020916A  RFA020917A  RFA022213A  RFA022393A  RFA022398A  RFA022400A  RFA022401A  RFA022416A  RFA022423A  RFA022426A  RFA022433A  RFA022435A  RFA022436A  RFA022439A  RFA022503A  RFA022508A  RFA022625A  RFA022650A  RFA022683A  RFA022687A  RFA022713A  RFA022727A  RFA022818A  RFA022820A  RFA022824A  RFA022825A  RFA022827A  RFA022828A  RFA022829A  RFA022831A  RFA022839A  RFA022841A  RFA022842A  RFA022846A",
      "center_classification_date": "20200705",
      "distribution_pattern": "worldwide",
      "state": "MN",
      "product_description": "Patient Connector, Model Number 24967",
      "report_date": "20200715",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-2506-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85816",
      "more_code_info": "",
      "recall_initiation_date": "20200604",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}