{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98994",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.",
      "recall_number": "Z-2479-2026",
      "product_description": "MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes:    1.\tDYND13508 - 08 Fr;  2.\tDYND13510 - 10 Fr;  3.\tDYND13512 - 12 Fr;  4.\tDYND13514 - 14 Fr;  5.\tDYND13515 - 15 Fr;  6.\tDYND13516 - 16 Fr;  7.\tDYND13518 - 18 Fr;  8.\tDYND13520 - 20 Fr;  9.\tDYND13522 - 22 Fr;  10.\tDYND13614 - 14 Fr;  11.\tDYND13616 - 16 Fr.",
      "product_quantity": "2237471 units",
      "reason_for_recall": "During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.",
      "recall_initiation_date": "20260504",
      "center_classification_date": "20260617",
      "report_date": "20260624",
      "code_info": "1. DYND13508, UDI/DI 10080196031139 (each), 20080196031136 (case);  2. DYND13510, UDI/DI 10080196726462 (each), 40080196726463 (case);  3. DYND13512, UDI/DI 10080196031177 (each), 40080196031178 (case);  4. DYND13514, UDI/DI 10080196726219 (each), 40080196726210 (case);  5. DYND13515, UDI/DI 10080196031214 (each), 40080196031215 (case);  6. DYND13516, UDI/DI 10080196031221 (each), 40080196031222 (case);  7. DYND13518, UDI/DI 10080196031245 (each), 40080196031246 (case);  8. DYND13520, UDI/DI 10080196031252 (each), 40080196031253 (case);  9. DYND13522, UDI/DI 10080196061259 (each), 40080196061250 (case);  10. DYND13614, UDI/DI 10080196031276 (each), 30080196031270 (case);  11. DYND13616, UDI/DI 10888277720879 (each), 20888277720876 (case);  ALL LOTS",
      "more_code_info": ""
    }
  ]
}