      "status": "Terminated",
      "city": "Athens",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88498",
      "recalling_firm": "Quidel Corporation",
      "address_1": "2005 E State St Ste 100",
      "address_2": "N/A",
      "postal_code": "45701-2125",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including Puerto Rico and the countries of Canada, Panama, Virgin Islands, Germany, Honduras.",
      "recall_number": "Z-2477-2021",
      "product_description": "Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization-  Ref: M124.    The Lyra Direct SARS-CoV-2 Assay is a real-time RT-PCR assay intended for the qualitative detection of nucleic acid from SARS-CoV-2 in nasal (NS), nasopharyngeal (NP), or oropharyngeal (OP) direct swab specimens from individuals suspected of COVID-19 by their healthcare provider",
      "product_quantity": "81,910 kits",
      "reason_for_recall": "Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold (Ct) value of 5 when using the Thermo Fisher QuantStudio 7 Pro, the Applied Biosystems 7500 Fast Dx, the Applied Biosystems 7500, the Bio-Rad CFX96 Touch\", the Roche LightCycler LC 480 II/ cobas z 480, or the Qiagen Rotor-Gene¿ MDx.",
      "recall_initiation_date": "20210813",
      "center_classification_date": "20210915",
      "termination_date": "20231102",
      "report_date": "20210922",
      "code_info": "All lots",
      "more_code_info": ""
    }
  ]
}