{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88546",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of ID, MA, MI, TX, VA.",
      "recall_number": "Z-2476-2021",
      "product_description": "The Azurion series (within the limits of the used Operation Room table) are intended for use to perform:  \"Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.  \"Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures.  Additionally:  \"The Azurion series can be used in a hybrid Operation Room.  \"The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.",
      "product_quantity": "1 unit",
      "reason_for_recall": "Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU.     The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails.    When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.",
      "recall_initiation_date": "20210813",
      "center_classification_date": "20210914",
      "termination_date": "20240311",
      "report_date": "20210922",
      "code_info": "Serial number: 921",
      "more_code_info": ""
    }
  ]
}