  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63091",
      "recalling_firm": "Alere San Diego, Inc.",
      "address_1": "9975 Summers Ridge Rd",
      "address_2": "N/A",
      "postal_code": "92121-2997",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-2460-2012",
      "product_description": "Alere Triage TOX Drug Screen 9 Panel, P/N: 94402.    The Alere Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses.",
      "product_quantity": "591 kits",
      "reason_for_recall": "Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.",
      "recall_initiation_date": "20120712",
      "center_classification_date": "20120926",
      "termination_date": "20130201",
      "report_date": "20121003",
      "code_info": "Lot Codes: K51245RB, W51022RB, W51207RB",
      "more_code_info": ""
    }
  ]
}