{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92808",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA, IA, MD, MI, NJ, NM, NY, OH, PA, SD, TX, UT, and VA. Global distribution.",
      "recall_number": "Z-2451-2023",
      "product_description": "ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151,  in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid",
      "product_quantity": "3067 units",
      "reason_for_recall": "There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.",
      "recall_initiation_date": "20230725",
      "center_classification_date": "20230824",
      "report_date": "20230830",
      "code_info": "UDI-DI: 00630414279176; Lot Numbers: All lots"
    }
  ]
}