  "results": [
    {
      "status": "Ongoing",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94690",
      "recalling_firm": "ICU Medical Inc",
      "address_1": "600 N Field Dr",
      "address_2": "",
      "postal_code": "60045-4835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay.",
      "recall_number": "Z-2449-2024",
      "product_description": "Plum 360/A+ spare batteries. Item Number: SUB0000864.",
      "product_quantity": "130,826 units",
      "reason_for_recall": "Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.",
      "recall_initiation_date": "20240507",
      "center_classification_date": "20240809",
      "report_date": "20240821",
      "code_info": "Item Number: SUB0000864; Lot Numbers: 13620902, 13588166, 13651642, 13651688, 13651686, 13651685, 13492940, 13498252, 13498259, 13716418, 13716411, 13651680, 13620909, 13716514, 13588168, 13594285, 13620897, 13638726, 13620893, 13594289, 13638724, 13651671, 13594281, 13620898, 13659496, 13638723, 13594280, 13638728, 13588163, 13588169, 13588170, 13651640, 13638727, 13618384, 13769263, 13588165, 13588161, 13716427, 13570187, 13765276, 13651678, 13746712, 13749803, 13771023, 13771500, 13716336, 13588159, 13716329, 13594286, 13594287, 13620905, 13651677, 13594282.",
      "more_code_info": ""
    }
  ]
}