      "status": "Terminated",
      "city": "Queensbury",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83334",
      "recalling_firm": "Angiodynamics, Inc.",
      "address_1": "603 Queensbury Ave",
      "address_2": "N/A",
      "postal_code": "12804-7619",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "The products were distributed to the following US states:  DC, IL, IN, MI, NJ, TN, TX, and WI.",
      "recall_number": "Z-2444-2019",
      "product_description": "AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012",
      "product_quantity": "109",
      "reason_for_recall": "The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.",
      "recall_initiation_date": "20190627",
      "center_classification_date": "20190830",
      "termination_date": "20200729",
      "report_date": "20190911",
      "code_info": "Lot 5456490; UPN H787114030120"
    }
  ]
}