  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97233",
      "recalling_firm": "VANTIVE US HEALTHCARE LLC",
      "address_1": "510 Lake Cook Rd",
      "address_2": "",
      "postal_code": "60015-4964",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-2437-2025",
      "product_description": "PrisMax V2-US, Hemodialysis Delivery System, Product Codes:  955626, 955558, 955627, 955701",
      "product_quantity": "4772 units",
      "reason_for_recall": "Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles",
      "recall_initiation_date": "20250728",
      "center_classification_date": "20250826",
      "report_date": "20250903",
      "code_info": "UDI 00085412639499, All Serial Numbers",
      "more_code_info": ""
    }
  ]
}