      "status": "Ongoing",
      "city": "Breinigsville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94890",
      "recalling_firm": "B Braun Medical Inc",
      "address_1": "200 Boulder Dr",
      "address_2": "",
      "postal_code": "18031-1532",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. International distribution to Canada.",
      "recall_number": "Z-2435-2024",
      "product_description": "Brand Name: Perfusor Space  Product Name: Perfusor PCA Syringe Pump  Model/Catalog Number: 8713080U  Product Description: Perfusor¿ Space Syringe Pump, Infusion Pump, 1 each  Component: N/A",
      "product_quantity": "1663",
      "reason_for_recall": "Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.",
      "recall_initiation_date": "20240614",
      "center_classification_date": "20240725",
      "report_date": "20240731",
      "code_info": "Lot Code: Model No 8713080U; UDI-DI 4046963915308; All Serial Numbers",
      "more_code_info": ""
    }
  ]
}