  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92561",
      "recalling_firm": "Carefusion 2200 Inc",
      "address_1": "75 N Fairway Dr",
      "address_2": "",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "US and Canada",
      "recall_number": "Z-2422-2023",
      "product_description": "Snowden-Pencer MIS  DIAMOND-TOUCH CLAMP BOWEL 60 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7046; endoscopic surgical device",
      "product_quantity": "4 units",
      "reason_for_recall": "Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.",
      "recall_initiation_date": "20230713",
      "center_classification_date": "20230817",
      "report_date": "20230823",
      "code_info": "UDI/DI 10885403160578, Lot Numbers:  C23, L22",
      "more_code_info": ""
    }
  ]
}