  "results": [
    {
      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage.  Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert.  Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.",
      "address_2": "",
      "product_quantity": "1867 units",
      "code_info": "Product UPN/GTIN:  00613994425010  Serial Numbers:  PNR611652S  PNR611653S  PNR611654S  PNR611655S  PNR611656S  PNR611657S  PNR611658S  PNR611659S  PNR611661S  PNR611662S  PNR611663S  PNR611664S  PNR611665S  PNR611666S  PNR611667S  PNR611669S  PNR611670S  PNR611671S  PNR611672S  PNR611673S  PNR611674S  PNR611675S  PNR611724S  PNR611726S  PNR611728S  PNR611730S  PNR611734S  PNR611735S  PNR611738S  PNR611743S  PNR611746S  PNR611749S  PNR611750S  PNR611751S  PNR611752S  PNR611753S  PNR611754S  PNR611755S  PNR611756S  PNR611757S  PNR611758S  PNR611759S  PNR611761S  PNR611762S  PNR611763S  PNR611764S  PNR611765S  PNR611767S  PNR611768S  PNR611861S  PNR611862S  PNR611863S  PNR611864S  PNR611865S  PNR611866S  PNR611867S  PNR611868S  PNR611869S  PNR611870S  PNR611871S  PNR611872S  PNR611873S  PNR611875S  PNR611876S  PNR611877S  PNR611878S  PNR611879S  PNR611880S  PNR611881S  PNR611883S  PNR611884S  PNR611885S  PNR611886S  PNR611888S  PNR611889S  PNR611891S  PNR611892S  PNR611893S  PNR611894S  PNR611895S  PNR611896S  PNR611897S  PNR611898S  PNR611899S  PNR611900S  PNR611901S  PNR611902S  PNR611903S  PNR611904S  PNR611905S  PNR611909S  PNR611916S  PNR611917S  PNR611929S  PNR611930S  PNR611931S  PNR611933S  PNR611934S  PNR611936S  PNR611937S  PNR611938S  PNR611939S  PNR611940S  PNR611941S  PNR611943S  PNR611944S  PNR611945S  PNR611947S  PNR611948S  PNR611955S  PNR611956S  PNR611957S  PNR611958S  PNR611959S  PNR611961S  PNR611962S  PNR611964S  PNR611978S  PNR611979S  PNR611980S  PNR611981S  PNR611982S  PNR611983S  PNR611984S  PNR611985S  PNR611986S  PNR611987S  PNR611988S  PNR611989S  PNR611990S  PNR611991S  PNR611992S  PNR611994S  PNR611995S  PNR611997S  PNR611998S  PNR611999S  PNR612000S  PNR612003S  PNR612004S  PNR612005S  PNR612006S  PNR612007S  PNR612008S  PNR612009S  PNR612010S  PNR612011S  PNR612012S  PNR612013S  PNR612014S  PNR612015S  PNR612016S  PNR612017S  PNR612018S  PNR612019S  PNR612020S  PNR612021S  PNR612022S  PNR612023S  PNR612024S  PNR612025S  PNR612026S  PNR612027S  PNR612028S  PNR612029S  PNR612030S  PNR612031S  PNR612032S  PNR612033S  PNR612034S  PNR612036S  PNR612037S  PNR612038S  PNR612039S  PNR612040S  PNR612042S  PNR612043S  PNR612044S  PNR612045S  PNR612047S  PNR612048S  PNR612050S  PNR612051S  PNR612052S  PNR612053S  PNR612054S  PNR612055S  PNR612056S  PNR612057S  PNR612058S  PNR612059S  PNR612060S  PNR612061S  PNR612062S  PNR612063S  PNR612064S  PNR612065S  PNR612066S  PNR612067S  PNR612068S  PNR612073S  PNR612075S  PNR612076S  PNR612077S  PNR612078S  PNR612079S  PNR612080S  PNR612081S  PNR612082S  PNR612083S  PNR612084S  PNR612085S  PNR612086S  PNR612087S  PNR612089S  PNR612090S  PNR612092S  PNR612093S  PNR612094S  PNR612095S  PNR612096S  PNR612097S  PNR612098S  PNR612099S  PNR612100S  PNR612101S  PNR612102S  PNR612103S  PNR612104S  PNR612105S  PNR612106S  PNR612107S  PNR612108S  PNR612109S  PNR612110S  PNR612111S  PNR612112S  PNR612113S  PNR612114S  PNR612115S  PNR612116S  PNR612117S  PNR612118S  PNR612119S  PNR612120S  PNR612121S  PNR612122S  PNR612123S  PNR612124S  PNR612125S  PNR612126S  PNR612127S  PNR612128S  PNR612129S  PNR612131S  PNR612132S  PNR612133S  PNR612134S  PNR612135S  PNR612136S  PNR612140S  PNR612141S  PNR612171S  PNR612173S  PNR612178S  PNR612180S  PNR612182S  PNR612188S  PNR612247S  PNR612252S  PNR612255S  PNR612424S  PNR612425S  PNR612427S  PNR612607S  PNR612608S  PNR612609S  PNR612610S  PNR612611S  PNR612612S  PNR612613S  PNR612614S  PNR612615S  PNR612616S  PNR612617S  PNR612618S  PNR612619S  PNR612620S  PNR612621S  PNR612799S  PNR612800S  PNR612801S  PNR612802S  PNR612803S  PNR612805S  PNR612806S  PNR612807S  PNR612808S  PNR612809S  PNR612810S  PNR612811S  PNR612812S  PNR612813S  PNR612814S  PNR612815S  PNR612816S  PNR612817S  PNR612818S  PNR612824S  PNR612825S  PNR612826S  PNR612827S  PNR612828S  PNR612829S  PNR612830S  PNR612831S  PNR612832S  PNR612834S  PNR612836S  PNR612837S  PNR612838S  PNR612839S  PNR612840S  PNR612841S  PNR612842S  PNR612844S  PNR612846S  PNR612847S  PNR612848S  PNR612849S  PNR612850S  PNR612851S  PNR612852S  PNR612865S  PNR612866S  PNR612867S  PNR612868S  PNR612869S  PNR612870S  PNR612871S  PNR612872S  PNR612873S  PNR612874S  PNR612876S  PNR612877S  PNR612878S  PNR612879S  PNR612905S  PNR612907S  PNR612908S  PNR612944S  PNR612945S  PNR612947S  PNR612948S  PNR612949S  PNR612950S  PNR612952S  PNR612953S  PNR612954S  PNR612956S  PNR612957S  PNR612958S  PNR612959S  PNR612960S  PNR612961S  PNR612962S  PNR612963S  PNR612964S  PNR612965S  PNR612966S  PNR612967S  PNR612968S  PNR612969S  PNR612970S  PNR612971S  PNR612972S  PNR612973S  PNR612974S  PNR612975S  PNR612976S  PNR612977S  PNR612978S  PNR612979S  PNR612985S  PNR612986S  PNR612987S  PNR612989S  PNR612991S  PNR613004S  PNR613006S  PNR613011S  PNR613013S  PNR613015S  PNR613016S  PNR613018S  PNR613020S  PNR613022S  PNR613023S  PNR613024S  PNR613025S  PNR613026S  PNR613072S  PNR613073S  PNR613074S  PNR613075S  PNR613077S  PNR613078S  PNR613079S  PNR613080S  PNR613081S  PNR613082S  PNR613083S  PNR613084S  PNR613085S  PNR613086S  PNR613117S  PNR613118S  PNR613125S  PNR613126S  PNR613127S  PNR613128S  PNR613129S  PNR613130S  PNR613131S  PNR613132S  PNR613134S  PNR613135S  PNR613136S  PNR613138S  PNR613140S  PNR613142S  PNR613143S  PNR613144S  PNR613145S  PNR613147S  PNR613188S  PNR613189S  PNR613190S  PNR613191S  PNR613192S  PNR613193S  PNR613194S  PNR613195S  PNR613196S  PNR613197S  PNR613198S  PNR613199S  PNR613200S  PNR613201S  PNR613202S  PNR613203S  PNR613205S  PNR613206S  PNR613207S  PNR613211S  PNR613212S  PNR613213S  PNR613215S  PNR613216S  PNR613217S  PNR613218S  PNR613219S  PNR613221S  PNR613222S  PNR613223S  PNR613225S  PNR613226S  PNR613227S  PNR613228S  PNR613230S  PNR613231S  PNR613233S  PNR613234S  PNR613235S  PNR613237S  PNR613238S  PNR613239S  PNR613240S  PNR613241S  PNR613242S  PNR613243S  PNR613244S  PNR613245S  PNR613247S  PNR613248S  PNR613249S  PNR613250S  PNR613251S  PNR613252S  PNR613253S  PNR613254S  PNR613255S  PNR613256S  PNR613257S  PNR613258S  PNR613259S  PNR613273S  PNR613275S  PNR613277S  PNR613282S  PNR613288S  PNR613290S  PNR613297S  PNR613298S  PNR613299S  PNR613300S  PNR613301S  PNR613302S  PNR613303S  PNR613304S  PNR613305S  PNR613306S  PNR613307S  PNR613308S  PNR613309S  PNR613310S  PNR613311S  PNR613312S  PNR613313S  PNR613314S  PNR613315S  PNR613316S  PNR613317S  PNR613318S  PNR613319S  PNR613320S  PNR613321S  PNR613322S  PNR613323S  PNR613325S  PNR613326S  PNR613327S  PNR613329S  PNR613330S  PNR613331S  PNR613332S  PNR613333S  PNR613334S  PNR613335S  PNR613336S  PNR613337S  PNR613338S  PNR613339S  PNR613340S  PNR613341S  PNR613342S  PNR613343S  PNR613344S  PNR613345S  PNR613346S  PNR613347S  PNR613348S  PNR613349S  PNR613350S  PNR613351S  PNR613352S  PNR613353S  PNR613354S  PNR613355S  PNR613356S  PNR613357S  PNR613358S  PNR613359S  PNR613360S  PNR613361S  PNR613363S  PNR613364S  PNR613365S  PNR613366S  PNR613367S  PNR613368S  PNR613369S  PNR613370S  PNR613371S  PNR613372S  PNR613373S  PNR613374S  PNR613375S  PNR613376S  PNR613377S  PNR613379S  PNR613380S  PNR613381S  PNR613382S  PNR613383S  PNR613384S  PNR613385S  PNR613386S  PNR613387S  PNR613388S  PNR613389S  PNR613390S  PNR613391S  PNR613392S  PNR613394S  PNR613395S  PNR613396S  PNR613397S  PNR613398S  PNR613399S  PNR613400S  PNR613401S  PNR613402S  PNR613403S  PNR613404S  PNR613405S  PNR613406S  PNR613407S  PNR613408S  PNR613410S  PNR613411S  PNR613412S  PNR613413S  PNR613414S  PNR613415S  PNR613416S  PNR613417S  PNR613418S  PNR613419S  PNR613420S  PNR613421S  PNR613422S  PNR613423S  PNR613424S  PNR613425S  PNR613426S  PNR613427S  PNR613428S  PNR613429S  PNR613430S  PNR613431S  PNR613432S  PNR613433S  PNR613434S  PNR613435S  PNR613436S  PNR613437S  PNR613438S  PNR613439S  PNR613441S  PNR613442S  PNR613443S  PNR613444S  PNR613445S  PNR613446S  PNR613447S  PNR613449S  PNR613450S  PNR613451S  PNR613452S  PNR613453S  PNR613454S  PNR613455S  PNR613456S  PNR613457S  PNR613458S  PNR613459S  PNR613460S  PNR613461S  PNR613462S  PNR613463S  PNR613464S  PNR613465S  PNR613466S  PNR613467S  PNR613468S  PNR613470S  PNR613471S  PNR613472S  PNR613473S  PNR613474S  PNR613475S  PNR613477S  PNR613478S  PNR613479S  PNR613480S  PNR613481S  PNR613482S  PNR613483S  PNR613484S  PNR613485S  PNR613486S  PNR613487S  PNR613488S  PNR613489S  PNR613490S  PNR613491S  PNR613492S  PNR613493S  PNR613494S  PNR613495S  PNR613496S  PNR613497S  PNR613498S  PNR613499S  PNR613500S  PNR613501S  PNR613502S  PNR613503S  PNR613504S  PNR613505S  PNR613506S  PNR613507S  PNR613508S  PNR613509S  PNR613510S  PNR613511S  PNR613512S  PNR613513S  PNR613514S  PNR613515S  PNR613516S  PNR613517S  PNR613518S  PNR613519S  PNR613520S  PNR613521S  PNR613522S  PNR613523S  PNR613524S  PNR613525S  PNR613526S  PNR613527S  PNR613528S  PNR613530S  PNR613531S  PNR613532S  PNR613533S  PNR613534S  PNR613535S  PNR613536S  PNR613537S  PNR613538S  PNR613539S  PNR613540S  PNR613541S  PNR613542S  PNR613543S  PNR613544S  PNR613545S  PNR613546S  PNR613547S  PNR613548S  PNR613549S  PNR613550S  PNR613551S  PNR613553S  PNR613554S  PNR613555S  PNR613556S  PNR613557S  PNR613558S  PNR613559S  PNR613560S  PNR613561S  PNR613562S  PNR613563S  PNR613564S  PNR613565S  PNR613566S  PNR613567S  PNR613568S  PNR613569S  PNR613570S  PNR613571S  PNR613572S  PNR613573S  PNR613574S  PNR613575S  PNR613576S  PNR613578S  PNR613579S  PNR613580S  PNR613581S  PNR613582S  PNR613583S  PNR613584S  PNR613585S  PNR613586S  PNR613587S  PNR613588S  PNR613589S  PNR613602S  PNR613603S  PNR613604S  PNR613606S  PNR613607S  PNR613608S  PNR613609S  PNR613610S  PNR613611S  PNR613612S  PNR613613S  PNR613614S  PNR613615S  PNR613616S  PNR613617S  PNR613618S  PNR613619S  PNR613620S  PNR613621S  PNR613622S  PNR613623S  PNR613624S  PNR613625S  PNR613627S  PNR613628S  PNR613630S  PNR613631S  PNR613632S  PNR613633S  PNR613634S  PNR613635S  PNR613636S  PNR613637S  PNR613638S  PNR613639S  PNR613640S  PNR613641S  PNR613642S  PNR613643S  PNR613644S  PNR613645S  PNR613646S  PNR613647S  PNR613648S  PNR613649S  PNR613650S  PNR613651S  PNR613652S  PNR613653S  PNR613654S  PNR613655S  PNR613656S  PNR613657S  PNR613658S  PNR613659S  PNR613660S  PNR613661S  PNR613662S  PNR613663S  PNR613664S  PNR613665S  PNR613666S  PNR613667S  PNR613668S  PNR613669S  PNR613670S  PNR613671S  PNR613672S  PNR613673S  PNR613674S  PNR613675S  PNR613676S  PNR613677S  PNR613678S  PNR613681S  PNR613682S  PNR613683S  PNR613684S  PNR613685S  PNR613686S  PNR613687S  PNR613688S  PNR613689S  PNR613691S  PNR613692S  PNR613693S  PNR613695S  PNR613696S  PNR613697S  PNR613698S  PNR613699S  PNR613700S  PNR613701S  PNR613702S  PNR613703S  PNR613704S  PNR613705S  PNR613706S  PNR613707S  PNR613708S  PNR613710S  PNR613711S  PNR613712S  PNR613715S  PNR613716S  PNR613717S  PNR613719S  PNR613721S  PNR613723S  PNR613725S  PNR613726S  PNR613727S  PNR613728S  PNR613729S  PNR613731S  PNR613733S  PNR613734S  PNR613735S  PNR613737S  PNR613738S  PNR613739S  PNR613740S  PNR613741S  PNR613742S  PNR613743S  PNR613744S  PNR613745S  PNR613746S  PNR613747S  PNR613748S  PNR613749S  PNR613751S  PNR613752S  PNR613753S  PNR613754S  PNR613755S  PNR613756S  PNR613757S  PNR613758S  PNR613759S  PNR613760S  PNR613761S  PNR613762S  PNR613763S  PNR613764S  PNR613765S  PNR613766S  PNR613767S  PNR613770S  PNR613771S  PNR613772S  PNR613773S  PNR613774S  PNR613775S  PNR613776S  PNR613779S  PNR613780S  PNR613781S  PNR613782S  PNR613783S  PNR613784S  PNR613785S  PNR613786S  PNR613787S  PNR613788S  PNR613790S  PNR613791S  PNR613792S  PNR613793S  PNR613794S  PNR613795S  PNR613796S  PNR613797S  PNR613798S  PNR613799S  PNR613800S  PNR613801S  PNR613802S  PNR613803S  PNR613804S  PNR613806S  PNR613807S  PNR613808S  PNR613809S  PNR613810S  PNR613811S  PNR613812S  PNR613813S  PNR613814S  PNR613815S  PNR613817S  PNR613819S  PNR613820S  PNR613821S  PNR613822S  PNR613823S  PNR613824S  PNR613825S  PNR613826S  PNR613827S  PNR613828S  PNR613829S  PNR613830S  PNR613831S  PNR613832S  PNR613833S  PNR613834S  PNR613835S  PNR613836S  PNR613837S  PNR613838S  PNR613839S  PNR613841S  PNR613842S  PNR613844S  PNR613845S  PNR613846S  PNR613847S  PNR613848S  PNR613849S  PNR613850S  PNR613851S  PNR613852S  PNR613853S  PNR613854S  PNR613855S  PNR613856S  PNR613857S  PNR613858S  PNR613859S  PNR613860S  PNR613861S  PNR613862S  PNR613863S  PNR613864S  PNR613865S  PNR613866S  PNR613867S  PNR613868S  PNR613869S  PNR613870S  PNR613871S  PNR613872S  PNR613873S  PNR613874S  PNR613875S  PNR613876S  PNR613877S  PNR613878S  PNR613879S  PNR613880S  PNR613881S  PNR613882S  PNR613883S  PNR613884S  PNR613885S  PNR613886S  PNR613887S  PNR613888S  PNR613889S  PNR613890S  PNR613891S  PNR613892S  PNR613893S  PNR613894S  PNR613895S  PNR613896S  PNR613897S  PNR613898S  PNR613899S  PNR613900S  PNR613901S  PNR613902S  PNR613903S  PNR613904S  PNR613905S  PNR613906S  PNR613907S  PNR613908S  PNR613909S  PNR613910S  PNR613911S  PNR613912S  PNR613913S  PNR613914S  PNR613915S  PNR613916S  PNR613917S  PNR613918S  PNR613919S  PNR613921S  PNR613922S  PNR613923S  PNR613924S  PNR613925S  PNR613926S  PNR613927S  PNR613928S  PNR613929S  PNR613930S  PNR613931S  PNR613932S  PNR613933S  PNR613934S  PNR613935S  PNR613936S  PNR613937S  PNR613938S  PNR613939S  PNR613940S  PNR613941S  PNR613942S  PNR613943S  PNR613944S  PNR613945S  PNR613946S  PNR613947S  PNR613948S  PNR613949S  PNR613950S  PNR613951S  PNR613952S  PNR613953S  PNR613954S  PNR613955S  PNR613956S  PNR613957S  PNR613958S  PNR613959S  PNR613960S  PNR613961S  PNR613962S  PNR613963S  PNR613964S  PNR613965S  PNR613966S  PNR613967S  PNR613968S  PNR613969S  PNR613970S  PNR613986S  PNR613987S  PNR613988S  PNR613989S  PNR613990S  PNR613991S  PNR613992S  PNR613993S  PNR613994S  PNR613995S  PNR613996S  PNR613997S  PNR613998S  PNR613999S  PNR614000S  PNR614001S  PNR614002S  PNR614003S  PNR614004S  PNR614005S  PNR614006S  PNR614008S  PNR614011S  PNR614012S  PNR614013S  PNR614014S  PNR614015S  PNR614016S  PNR614017S  PNR614018S  PNR614019S  PNR614020S  PNR614021S  PNR614022S  PNR614023S  PNR614024S  PNR614025S  PNR614026S  PNR614027S  PNR614028S  PNR614029S  PNR614030S  PNR614031S  PNR614032S  PNR614033S  PNR614034S  PNR614035S  PNR614036S  PNR614037S  PNR614038S  PNR614039S  PNR614040S  PNR614041S  PNR614042S  PNR614043S  PNR614044S  PNR614045S  PNR614046S  PNR614047S  PNR614048S  PNR614049S  PNR614050S  PNR614051S  PNR614052S  PNR614053S  PNR614054S  PNR614055S  PNR614056S  PNR614057S  PNR614058S  PNR614059S  PNR614060S  PNR614061S  PNR614062S  PNR614063S  PNR614064S  PNR614065S  PNR614066S  PNR614068S  PNR614069S  PNR614070S  PNR614071S  PNR614072S  PNR614073S  PNR614074S  PNR614076S  PNR614077S  PNR614078S  PNR614079S  PNR614080S  PNR614081S  PNR614082S  PNR614083S  PNR614084S  PNR614085S  PNR614086S  PNR614087S  PNR614088S  PNR614090S  PNR614091S  PNR614092S  PNR614093S  PNR614094S  PNR614095S  PNR614096S  PNR614098S  PNR614099S  PNR614100S  PNR614101S  PNR614102S  PNR614103S  PNR614104S  PNR614105S  PNR614107S  PNR614108S  PNR614109S  PNR614110S  PNR614111S  PNR614112S  PNR614113S  PNR614114S  PNR614115S  PNR614116S  PNR614117S  PNR614118S  PNR614119S  PNR614120S  PNR614121S  PNR614122S  PNR614124S  PNR614126S  PNR614127S  PNR614128S  PNR614129S  PNR614130S  PNR614131S  PNR614132S  PNR614133S  PNR614134S  PNR614136S  PNR614137S  PNR614138S  PNR614139S  PNR614141S  PNR614142S  PNR614143S  PNR614160S  PNR614161S  PNR614163S  PNR614164S  PNR614165S  PNR614166S  PNR614167S  PNR614168S  PNR614169S  PNR614170S  PNR614171S  PNR614174S  PNR614175S  PNR614176S  PNR614177S  PNR614178S  PNR614179S  PNR614180S  PNR614184S  PNR614185S  PNR614186S  PNR614187S  PNR614188S  PNR614189S  PNR614190S  PNR614191S  PNR614192S  PNR614193S  PNR614194S  PNR614195S  PNR614196S  PNR614197S  PNR614198S  PNR614199S  PNR614200S  PNR614201S  PNR614202S  PNR614203S  PNR614204S  PNR614205S  PNR614206S  PNR614207S  PNR614208S  PNR614209S  PNR614210S  PNR614211S  PNR614212S  PNR614213S  PNR614214S  PNR614215S  PNR614216S  PNR614217S  PNR614218S  PNR614219S  PNR614220S  PNR614221S  PNR614222S  PNR614223S  PNR614224S  PNR614225S  PNR614226S  PNR614227S  PNR614228S  PNR614229S  PNR614231S  PNR614232S  PNR614233S  PNR614234S  PNR614235S  PNR614236S  PNR614237S  PNR614239S  PNR614240S  PNR614241S  PNR614242S  PNR614243S  PNR614244S  PNR614245S  PNR614246S  PNR614247S  PNR614248S  PNR614249S  PNR614250S  PNR614251S  PNR614252S  PNR614253S  PNR614254S  PNR614256S  PNR614257S  PNR614258S  PNR614259S  PNR614260S  PNR614262S  PNR614264S  PNR614265S  PNR614266S  PNR614267S  PNR614268S  PNR614269S  PNR614270S  PNR614271S  PNR614272S  PNR614273S  PNR614274S  PNR614275S  PNR614276S  PNR614277S  PNR614278S  PNR614279S  PNR614280S  PNR614281S  PNR614282S  PNR614283S  PNR614284S  PNR614285S  PNR614286S  PNR614287S  PNR614288S  PNR614289S  PNR614290S  PNR614291S  PNR614292S  PNR614293S  PNR614294S  PNR614295S  PNR614297S  PNR614298S  PNR614299S  PNR614300S  PNR614301S  PNR614303S  PNR614304S  PNR614305S  PNR614307S  PNR614308S  PNR614309S  PNR614370S  PNR614371S  PNR614372S  PNR614373S  PNR614374S  PNR614375S  PNR614376S  PNR614377S  PNR614378S  PNR614379S  PNR614380S  PNR614381S  PNR614382S  PNR614384S  PNR614385S  PNR614386S  PNR614387S  PNR614389S  PNR614390S  PNR614391S  PNR614392S  PNR614393S  PNR614394S  PNR614395S  PNR614396S  PNR614397S  PNR614398S  PNR614399S  PNR614400S  PNR614401S  PNR614405S  PNR614407S  PNR614415S  PNR614416S  PNR614418S  PNR614419S  PNR614420S  PNR614421S  PNR614422S  PNR614423S  PNR614424S  PNR614425S  PNR614426S  PNR614427S  PNR614428S  PNR614429S  PNR614430S  PNR614431S  PNR614432S  PNR614433S  PNR614434S  PNR614435S  PNR614436S  PNR614437S  PNR614438S  PNR614439S  PNR614440S  PNR614441S  PNR614442S  PNR614443S  PNR614444S  PNR614445S  PNR614446S  PNR614447S  PNR614448S  PNR614449S  PNR614450S  PNR614451S  PNR614452S  PNR614453S  PNR614455S  PNR614456S  PNR614457S  PNR614458S  PNR614459S  PNR614460S  PNR614461S  PNR614463S  PNR614464S  PNR614465S  PNR614466S  PNR614467S  PNR614468S  PNR614470S  PNR614471S  PNR614472S  PNR614473S  PNR614475S  PNR614476S  PNR614477S  PNR614478S  PNR614479S  PNR614480S  PNR614481S  PNR614482S  PNR614483S  PNR614484S  PNR614485S  PNR614486S  PNR614487S  PNR614488S  PNR614489S  PNR614490S  PNR614491S  PNR614492S  PNR614494S  PNR614495S  PNR614496S  PNR614497S  PNR614498S  PNR614499S  PNR614500S  PNR614501S  PNR614502S  PNR614503S  PNR614504S  PNR614505S  PNR614506S  PNR614507S  PNR614508S  PNR614509S  PNR614510S  PNR614511S  PNR614512S  PNR614514S  PNR614515S  PNR614516S  PNR614517S  PNR614518S  PNR614519S  PNR614520S  PNR614521S  PNR614522S  PNR614523S  PNR614524S  PNR614525S  PNR614527S  PNR614528S  PNR614529S  PNR614530S  PNR614531S  PNR614533S  PNR614534S  PNR614535S  PNR614538S  PNR614539S  PNR614540S  PNR614541S  PNR614542S  PNR614543S  PNR614544S  PNR614545S  PNR614546S  PNR614547S  PNR614548S  PNR614549S  PNR614550S  PNR614551S  PNR614552S  PNR614553S  PNR614554S  PNR614555S  PNR614556S  PNR614557S  PNR614558S  PNR614560S  PNR614561S  PNR614562S  PNR614563S  PNR614564S  PNR614565S  PNR614566S  PNR614567S  PNR614568S  PNR614569S  PNR614571S  PNR614574S  PNR614575S  PNR614576S  PNR614578S  PNR614579S  PNR614582S  PNR614583S  PNR614584S  PNR614585S  PNR614586S  PNR614588S  PNR614589S  PNR614590S  PNR614591S  PNR614592S  PNR614593S  PNR614594S  PNR614595S  PNR614596S  PNR614597S  PNR614598S  PNR614599S  PNR614600S  PNR614601S  PNR614603S  PNR614604S  PNR614605S  PNR614606S  PNR614607S  PNR614609S  PNR614610S  PNR614611S  PNR614612S  PNR614613S  PNR614615S  PNR614616S  PNR614618S  PNR614621S  PNR614622S  PNR614624S  PNR614625S  PNR614627S  PNR614629S  PNR614631S  PNR614632S  PNR614633S  PNR614635S  PNR614636S  PNR614637S  PNR614638S  PNR614639S  PNR614640S  PNR614641S  PNR614642S  PNR614643S  PNR614644S  PNR614645S  PNR614646S  PNR614647S  PNR614648S  PNR614649S  PNR614650S  PNR614651S  PNR614652S  PNR614653S  PNR614654S  PNR614655S  PNR614656S  PNR614657S  PNR614659S  PNR614660S  PNR614661S  PNR614662S  PNR614663S  PNR614664S  PNR614665S  PNR614668S  PNR614670S  PNR614671S  PNR614672S  PNR614673S  PNR614674S  PNR614675S  PNR614676S  PNR614677S  PNR614678S  PNR614679S  PNR614680S  PNR614681S  PNR614685S  PNR614687S  PNR614688S  PNR614689S  PNR614690S  PNR614692S  PNR614693S  PNR614694S  PNR614697S  PNR614698S  PNR614700S  PNR614701S  PNR614702S  PNR614704S  PNR614705S  PNR614706S  PNR614709S  PNR614710S  PNR614711S  PNR614712S  PNR614713S  PNR614714S  PNR614715S  PNR614716S  PNR614717S  PNR614718S  PNR614720S  PNR614722S  PNR614723S  PNR614724S  PNR614727S  PNR614728S  PNR614729S  PNR614731S  PNR614732S  PNR614733S  PNR614734S  PNR614735S  PNR614736S  PNR614737S  PNR614738S  PNR614739S  PNR614740S  PNR614741S  PNR614742S  PNR614743S  PNR614744S  PNR614745S  PNR614746S  PNR614747S  PNR614748S  PNR614749S  PNR614750S  PNR614751S  PNR614752S  PNR614753S  PNR614754S  PNR614755S  PNR614756S  PNR614757S  PNR614758S  PNR614759S  PNR614760S  PNR614761S  PNR614764S  PNR614765S  PNR614766S  PNR614767S  PNR614768S  PNR614769S  PNR614770S  PNR614772S  PNR614773S  PNR614774S  PNR614775S  PNR614776S  PNR614777S  PNR614778S  PNR614779S  PNR614780S  PNR614781S  PNR614782S  PNR614783S  PNR614784S  PNR614785S  PNR614786S  PNR614787S  PNR614788S  PNR614792S  PNR614793S  PNR614795S  PNR614796S  PNR614797S  PNR614798S  PNR614799S  PNR614800S  PNR614801S  PNR614802S  PNR614804S  PNR614805S  PNR614806S  PNR614808S  PNR614809S  PNR614811S  PNR614812S  PNR614813S  PNR614814S  PNR614815S  PNR614816S  PNR614817S  PNR614818S  PNR614819S  PNR614820S  PNR614821S  PNR614822S  PNR614823S  PNR614824S  PNR614826S  PNR614827S  PNR614828S  PNR614829S  PNR614830S  PNR614831S  PNR614833S  PNR614834S  PNR614835S  PNR614837S  PNR614838S  PNR614839S  PNR614841S  PNR614842S  PNR614843S  PNR614844S  PNR614847S  PNR614848S  PNR614849S  PNR614850S  PNR614851S  PNR614852S  PNR614853S  PNR614854S  PNR614855S  PNR614856S  PNR614857S  PNR614858S  PNR614859S  PNR614861S  PNR614862S  PNR614864S  PNR614865S  PNR614866S  PNR614867S  PNR614868S  PNR614869S  PNR614870S  PNR614872S  PNR614873S  PNR614874S  PNR614875S  PNR614876S  PNR614877S  PNR614878S  PNR614879S  PNR614882S  PNR614883S  PNR614884S  PNR614885S  PNR614886S  PNR614898S  PNR614901S  PNR614902S  PNR614903S  PNR614904S  PNR614905S  PNR614906S  PNR614908S  PNR614910S  PNR614912S  PNR614913S  PNR614915S  PNR614916S  PNR614917S  PNR614918S  PNR614919S  PNR614920S  PNR614921S  PNR614922S  PNR614923S  PNR614924S  PNR614925S  PNR614926S  PNR614927S  PNR614928S  PNR614929S  PNR614930S  PNR614931S  PNR614932S  PNR615003S",
      "center_classification_date": "20180710",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "state": "MN",
      "product_description": "Medtronic EnTrust Escudo D144DRG (OUS distribution)    Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.",
      "report_date": "20180718",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-2421-2018",
      "initial_firm_notification": "Visit",
      "product_type": "Devices",
      "event_id": "80379",
      "termination_date": "20210616",
      "more_code_info": "",
      "recall_initiation_date": "20180621",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}