  "results": [
    {
      "status": "Ongoing",
      "city": "Rastatt",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97313",
      "recalling_firm": "MAQUET GMBH",
      "address_1": "Kehler Str. 31",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of  Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.",
      "recall_number": "Z-2420-2025",
      "product_description": "Corin Operating Tables  Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2;  All Software Versions.",
      "product_quantity": "315 units (8 US, 307 OUS)",
      "reason_for_recall": "Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.",
      "recall_initiation_date": "20250714",
      "center_classification_date": "20250822",
      "report_date": "20250903",
      "code_info": "Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; UDI-DI: (1) 04046768145177, (2) 04046768145184, (3) 04046768145191, (4) 04046768145207, (5) 04046768145214, (6) 04046768145221;  Lot/Serial Numbers: (1) 1, 2, 3, 10, 25, 26, 27, 28, 22, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 24,          (2) 9, 51, 53, 14, 15, 16, 17, 34, 35, 36, 37, 39, 40, 42, 43, 54, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 38, 44, 45, 46, 5, 13, 57, 58, 59, 18, 49, 47, 48, 7, 52,           (3) 41, 5, 6, 45, 48, 49, 50, 23, 38, 10, 12, 13, 14, 84, 87, 88, 89, 7, 63, 80, 82, 83, 86, 54, 93, 94, 11, 15, 18, 20, 21, 22, 27, 28, 32, 40, 9, 69, 70, 71, 72, 73, 74, 75, 76, 77, 78, 79, 85, 55, 56, 57, 58, 59, 42, 97, 92, 17, 26, 16, 24, 25, 30, 33, 35, 36, 37, 39, 44, 53, 91, 8, 19, 90, 46, 47, 51, 52, 34, 43, 60, 62, 64, 65, 66, 67, 68,           (4) 44, 70, 71, 42, 56, 65, 115, 185, 52, 105, 106, 107, 109, 145, 146, 147, 148, 11, 26, 19, 32, 33, 35, 36, 39, 53, 61, 63, 69, 14, 20, 68, 13, 18, 16, 101, 167, 170, 83, 3, 4, 23, 28, 37, 38, 40, 41, 43, 45, 47, 48, 50, 55, 58, 60, 62, 78, 81, 82, 92, 94, 96, 97, 98, 99, 100, 102, 116, 138, 139, 144, 150, 152, 153, 154, 155, 156, 165, 169, 171, 21, 22, 24, 25, 6, 17, 27, 29, 30, 34, 49, 51, 57, 64, 74, 75, 76, 77, 79, 80, 84, 85, 86, 87, 88, 89, 90, 91, 93, 95, 103, 104, 108, 110, 111, 112, 113, 114, 118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, 130, 131, 132, 133, 134, 135, 136, 137, 140, 141, 142, 143, 172, 173, 174, 194,           (5) 1, 2, 3, 4, 5, 6, 7,          (6) 12, 15;",
      "more_code_info": ""
    }
  ]
}