  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Reading",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85684",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "address_2": "N/A",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2417-2020",
      "product_description": "Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701",
      "product_quantity": "2123 total",
      "reason_for_recall": "A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate.  A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for  System Error 3  and  High Baseline  presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.",
      "recall_initiation_date": "20200520",
      "center_classification_date": "20200626",
      "termination_date": "20230107",
      "report_date": "20200708",
      "code_info": "Serial Number Ranges:   AC3 Optimus IABP NA/AJLA IAP-0701 170831F through 170840F     171011F through 171020F     171121F through 171125F     180106F through 180110F     180126F through 180130F     180426F through 180430F     180606F through 180610F     180731F through 180740F     180826F through 180830F     180901F through 180910F     180912F through 180921F     181001F through 181010F     190136F through 190140F     190336F through 190340F     190646F through 190650F     190731F through 190736F     190931F through 190936F     191019F through 191024F     191055F through 191060F     200107F through 200112F     200308F through 200311F"
    }
  ]
}