{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Amersham",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97297",
      "recalling_firm": "THOR Photomedicine Ltd",
      "address_1": "2 Anglo Office Park",
      "address_2": "Units 1 &; 67 White Lion Road",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.",
      "recall_number": "Z-2416-2025",
      "product_description": "Brand Name: THOR  Product Name: NovoTHOR  Model/Catalog Number: S2188  Software Version: N/A  Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles.  Component: No",
      "product_quantity": "80 units (66 US, 14 OUS)",
      "reason_for_recall": "The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.",
      "recall_initiation_date": "20250724",
      "center_classification_date": "20250822",
      "report_date": "20250903",
      "code_info": "Model/Catalog Number: S2188; UDI-DI: 05060494130503; Serial Numbers: 4004, 4015, 4017, 4019, 4021, 4022, 4024, 4026, 4028, 4029, 4035, 4036, 4037, 4042, 4043, 4044, 4046, 4047, 4049, 4054, 4055, 4059, 4066, 4071, 4072, 4073, 4074, 4076, 4077, 4078, 4079, 4082, 4083, 4084, 4085, 4086, 4087, 4088, 4090, 4091, 4094, 4095, 4097, 4099, 4101, 4105, 4106, 4111, 4112, 4114, 4116, 4117, 4121, 4123, 4124, 4125, 4126, 4130, 4140, 4141, 4143, 4144, 4151, 4153, 4157, 4158, 4009, 4020, 4103, 4107, 4109, 4110, 4113, 4120, 4122, 4127, 4133, 4134, 4138, 4139;",
      "more_code_info": ""
    }
  ]
}