  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Reading",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85684",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "address_2": "N/A",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2413-2020",
      "product_description": "Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J",
      "product_quantity": "2123 total",
      "reason_for_recall": "A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate.  A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for  System Error 3  and  High Baseline  presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.",
      "recall_initiation_date": "20200520",
      "center_classification_date": "20200626",
      "termination_date": "20230107",
      "report_date": "20200708",
      "code_info": "Serial Number Ranges:   AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W     150117W through 150118W     150120W     150201W     150305W     150307W through 150309W     150623W through 150624W     150713W through 150716W     151114W     151209W     151213W     151219W     160503W   160511W through 160515W      160517W     160635W through 160636W     160640W through 160641W     160731W through 160733W     171105W     171117W     171119W through 171120W     171201W     180120W     180301W"
    }
  ]
}