{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85687",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution including to the following states: Florida, New Jersey, New York, California, Arizona, Colorado, Pennsylvania, Virginia, Ohio, Texas, Alabama, Nebraska, Wisconsin, Montana, Kentucky, Washington DC, Massachusetts, North Carolina, Georgia, Louisiana, Kansas, South Carolina, Tennessee, Iowa, Rhode Island, South Dakota, New Hampshire, Maryland, Oregon, Indiana, Idaho, Connecticut, Puerto Rico, West Virginia, Washington, Illinois, Utah, Maine, Minnesota, Wyoming, Nevada, Michigan, Missouri, and New Mexico.",
      "recall_number": "Z-2405-2020",
      "product_description": "Atellica Solution IM1600 Analyzer, Siemens Material Number 11066000, UDI # 00630414002026",
      "product_quantity": "1329",
      "reason_for_recall": "Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and Atellica IM1600 analyzers.",
      "recall_initiation_date": "20200507",
      "center_classification_date": "20200619",
      "report_date": "20200701",
      "code_info": "Atellica Tube Top Sample Cups, 1 mL, SMN 11069061;    Atellica Tube Top Sample Cups, 2 mL, SMN 11069062;    Sarstedt 62.612, 15.3 x 92 False Bottom, 4 mL"
    }
  ]
}