  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mentor",
      "address_1": "5960 Heisley Rd",
      "reason_for_recall": "Customers may be storing objects on the base and/or around the column of the surgical table.  This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover.  Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .",
      "address_2": "",
      "product_quantity": "11,044 Units",
      "code_info": "B420702032 to 0402110097",
      "center_classification_date": "20120917",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including District of Columbia and  Puerto Rico and the following countries: American Samoa, Argentina, Australia, Bahrain, Bangladesh, Bermuda, Brazil, Canada, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, Germany, Greece, Guam, Hong Kong, India, Indonesia, Israel, Italy, Japa, Jordan, Kazakhstan, Korea, Kuwait, Lebanon , Libya , Malaysia , Mexico, Morocco, New Zealand, Panama, Peru, Philippines , Poland, Portugal, Qatar, Russia, Saudi Arabia , Singapore, South Africa, Spain, Sri Lanka, Taiwan, Thailand, Trinidad and Tobago, Turkey & United Kingdom.",
      "state": "OH",
      "product_description": "A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table.      Product Usage:  Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.",
      "report_date": "20120926",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Steris Corporation",
      "recall_number": "Z-2402-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "54705",
      "termination_date": "20130313",
      "more_code_info": "",
      "recall_initiation_date": "20100217",
      "postal_code": "44060",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}