{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mountain View",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83398",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "685 E Middlefield Rd",
      "address_2": "N/A",
      "postal_code": "94043-4045",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia,  Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.",
      "recall_number": "Z-2351-2019",
      "product_description": "ACUSON NX2 Elite Diagnostic Ultrasound System, Model # 11284500, RX Only, Siemens Medical Solutions, containing  software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869050775    The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.",
      "product_quantity": "70 systems",
      "reason_for_recall": "The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware",
      "recall_initiation_date": "20190128",
      "center_classification_date": "20190822",
      "report_date": "20190828",
      "code_info": "All product lots of the Acuson NX2, model # 11284500 containing the following software: VA10A, VA10B,VA10C, VA11A, VA11B, VA11C"
    }
  ]
}