  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Phoenix",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88331",
      "recalling_firm": "W. L. Gore & Associates Inc.",
      "address_1": "32360 N North Valley Pkwy",
      "address_2": "N/A",
      "postal_code": "85085-4228",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AUS, CAN, CHE, DEU, DNK, ESP, FIN, FRA, GBR, GRC, HRV, HUN, IRL, ISR, ITA, JPN, MEX, NLD, NOR, POL, PRT, SMR, SWE.",
      "recall_number": "Z-2340-2021",
      "product_description": "Gore Molding & Occlusion Balloon Catheter, REF Catalogue Number MOB37, SN XXXXXXXX, 10 mm - 37 m 10 Fr, Guidewire Compatibility 0.035\", STERILE EO, CE2797, UDI: (00)733132639489",
      "product_quantity": "3,364 units",
      "reason_for_recall": "Complaints received concerning Balloon Catheter leakage from the guidewire lumen and the y-hub during preparation resulting in either insufficient inflation or an inability to inflate.",
      "recall_initiation_date": "20210712",
      "center_classification_date": "20210825",
      "report_date": "20210901",
      "code_info": "Serial Numbers between 22982051 - 23516741"
    }
  ]
}