      "status": "Ongoing",
      "city": "Redwood City",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88337",
      "recalling_firm": "PROCEPT BIOROBOTICS CORPORATION",
      "address_1": "900 Island Dr Ste 101",
      "address_2": "N/A",
      "postal_code": "94065-5176",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, WI, AZ, AR, MI, WV, SD, PA, CA, DC, NJ, NY, GA, IN, VA, TN and the countries of Austria, Switzerland, Germany, Spain, France, UK, Italy.",
      "recall_number": "Z-2330-2021",
      "product_description": "AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.",
      "product_quantity": "608",
      "reason_for_recall": "Scope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both tip detachment and entire tip separation form the handpiece may lead to procedural delay for handpiece exchange. Tip detachment can be seen cystoscopically and tip separation from the handpiece may necessitate tip removal from the bladder or prostate.",
      "recall_initiation_date": "20210709",
      "center_classification_date": "20210820",
      "report_date": "20210901",
      "code_info": "Lot:  21C00226; 21C00290; 21C00304; 21C00359; 21C00373; 21C00464; 21C00465; 21C00527"
    }
  ]
}