  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88422",
      "recalling_firm": "Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states AR, CO, MN, MS, NJ, OK, PA, TX and the countries of China and Italy.",
      "recall_number": "Z-2329-2021",
      "product_description": "Triathlon Prim CEM FXD BPLT 7, Catalog No. 5520B700",
      "product_quantity": "14 US; 2 OUS",
      "reason_for_recall": "The internal profile of the Triathlon Prim (Baseplate) is potentially oversized; there may be a witness mark (taper) at the bottom of the internal profile of the baseplate, which prevents the insert from locking on the baseplate.",
      "recall_initiation_date": "20210701",
      "center_classification_date": "20210819",
      "termination_date": "20230517",
      "report_date": "20210825",
      "code_info": "Lot JHS2H"
    }
  ]
}