{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98880",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2316-2026",
      "product_description": "Allia IGS 7 Pulse angiographic X-ray system",
      "product_quantity": "35 units",
      "reason_for_recall": "GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.",
      "recall_initiation_date": "20260420",
      "center_classification_date": "20260605",
      "report_date": "20260617",
      "code_info": "UDI-DI: 00195278719720; Serial Numbers: D3-23-024, D3-23-029, D3-24-001, D3-24-002, D3-24-005, D3-24-006, D3-24-008, D3-24-009, D3-24-010, D3-24-013, D3-24-014, D3-24-016, D3-24-019, D3-24-020, D3-24-021, D3-24-023, D3-24-025, D3-25-001, D3-25-002, D3-25-003, D3-25-005, D3-25-006, D3-25-007, D3-25-008, D3-25-009, D3-25-010, D3-25-013, D3-25-016, D3-25-017, D3-25-018, D3-25-026, D3-25-027, D3-25-029, D3-25-030, D3-25-034, D3-25-036, D3-25-037, D3-26-001, D3-26-002, D3-26-007"
    }
  ]
}