  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98880",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2315-2026",
      "product_description": "Allia IGS 5 Pulse angiographic X-ray system",
      "product_quantity": "121 units",
      "reason_for_recall": "GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.",
      "recall_initiation_date": "20260420",
      "center_classification_date": "20260605",
      "report_date": "20260617",
      "code_info": "UDI-DI: 00195278719263; Serial Numbers: M2-23-039, M2-23-056, M2-24-004, M2-24-019, M2-24-020, M2-24-025, M2-24-030, M2-24-031, M2-24-032, M2-24-033, M2-24-034, M2-24-037, M2-24-038, M2-24-039, M2-24-040, M2-24-046, M2-24-057, M2-24-062, M2-24-064, M2-24-066, M2-24-069, M2-24-071, M2-24-077, M2-24-078, M2-24-080, M2-24-083, M2-25-003, M2-25-004, M2-25-005, M2-25-011, M2-25-013, M2-25-015, M2-25-017, M2-25-019, M2-25-023, M2-25-024, M2-25-025, M2-25-026, M2-25-028, M2-25-032, M2-25-036, M2-25-037, M2-25-039, M2-25-040, M2-25-041, M2-25-042, M2-25-043, M2-25-046, M2-25-048, M2-25-049, M2-25-056, M2-25-057, M2-25-058, M2-25-062, M2-25-065, M2-25-067, M2-25-072, M2-25-074, M2-25-077, M2-25-078, M2-25-080, M2-25-250, M2-25-300, M2-26-002, M2-26-004, M2-26-006, M2-26-007, M2-UP2502, M3-22-012, M3-23-117, M3-24-019, M3-24-021, M3-24-023, M3-24-034, M3-24-052, M3-24-061, M3-24-074, M3-24-079, M3-24-086, M3-24-090, M3-24-094, M3-24-095, M3-24-097, M3-24-099, M3-24-101, M3-24-104, M3-24-111, M3-24-120, M3-24-128, M3-24-136, M3-24-137, M3-24-141, M3-24-143, M3-24-149, M3-24-151, M3-24-156, M3-24-159, M3-24-161, M3-24-163, M3-24-165, M3-24-168, M3-24-169, M3-24-172, M3-24-175, M3-25-003, M3-25-012, M3-25-017, M3-25-020, M3-25-027, M3-25-036, M3-25-061, M3-25-063, M3-25-069, M3-25-074, M3-25-076, M3-25-079, M3-25-080, M3-25-084, M3-25-085, M3-25-088, M3-25-089, M3-25-094, M3-25-104, M3-25-106, M3-25-107, M3-25-115, M3-25-118, M3-25-119, M3-25-120, M3-25-128, M3-25-130, M3-25-132, M3-25-145, M3-25-147, M3-25-148, M3-25-149, M3-25-150, M3-25-151, M3-26-002, M3-26-003, M3-26-005, M3-26-007, M3-26-021, M3-26-024, M3-26-027, M3-26-028, M3-26-033, M3-26-034, M3-UP2501, M3-UP2503"
    }
  ]
}