  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83461",
      "recalling_firm": "Teleflex Medical",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2313-2019",
      "product_description": "Hudson RCI Sheridan EZ-ENDO:  a) 5.0 mm, REF 5-22510  b) 5.5 mm, REF 5-22511  c) 7.0 mm, REF 5-22514  d) 7.5 mm, REF 5-22515  e) 8.0 mm, REF 5-22516  f) 8.5 mm, REF 5-22517  g) 9.0 mm, REF 5-22518  h) 6.0 mm, REF 5-22512  i) 6.5 mm, REF 5-22513      Product Usage:  Tracheal tube/airway management",
      "product_quantity": "57181units",
      "reason_for_recall": "Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.",
      "recall_initiation_date": "20190730",
      "center_classification_date": "20190909",
      "termination_date": "20220215",
      "report_date": "20190918",
      "code_info": "a) REF 5-22510  Batch numbers:    73B1700556  73D1700650  73H1700346  73H1800449    b) REF 5-22511  Batch numbers:    73C1700261  73K1700609  73C1800370    c) REF 5-22514  Batch numbers:    73A1700118  73C1700569  73E1700722  73F1700175  73F1700405  73G1700433  73H1700349  73H1700494  73H1700711  73J1700143  73K1700081  73K1700610  73L1700181  73M1700273  73D1800653  73F1800562  73F1800720  73H1800450  73J1800531  73K1800328  73K1800786    d) REF 5-22515  Batch numbers:    73B1700316  73C1700262  73C1700418  73E1700723  73E1700104  73F1700176  73F1700406  73G1700434  73H1700350  73J1700302  73J1700497  73K1700747  73M1700274  73A1800068  73A1800683  73B1800502  73C1800201  73D1800154  73E1800402  73F1800565  73G1800591  73H1800451  73H1800686  73J1800113  73K1800736    e) REF 5-22516  Batch numbers:    73K1600594  73A1700282  73B1700158  73B1700317  73C1700570  73F1700036  73F1700177  73F1700407  73G1700614  73H1700495  73J1700487  73K1700082  73A1800684  73B1800503  73C1800371  73E1800577  73G1800232  73J1800532  73K1800327    f) REF 5-22517  Batch numbers:    73J1600767  73C1800202    g) REF 5-22518  Batch numbers:    73K1600595  73E1800459  73M1800288    h) REF 5-22512  Batch numbers:   73K1600152  73M1600163  73A1700280  73C1700567  73E1700496  73F1700403  73H1700347  73M1700272  73E1800401  73E1800575  73F1800354    i) REF 5-22513  Batch numbers:  73J1600766  73B1700155  73C1700568  73F1700404  73G1700432  73H1700348  73E1800576  73F1800355",
      "more_code_info": ""
    }
  ]
}