{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tiffin",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94765",
      "recalling_firm": "American Contract Systems Inc",
      "address_1": "85 Shaffer Park Dr",
      "address_2": "N/A",
      "postal_code": "44883-9290",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2297-2024",
      "product_description": "TOTAL KNEE - 297835 -Procedure Kit     Catalog Number: RGTK10L",
      "product_quantity": "24 units",
      "reason_for_recall": "Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.",
      "recall_initiation_date": "20240515",
      "center_classification_date": "20240705",
      "report_date": "20240717",
      "code_info": "UDI-DI: 00191072216182  LOT#\"s:  939241"
    }
  ]
}