{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "South Jordan",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97151",
      "recalling_firm": "Merit Medical Systems, Inc.",
      "address_1": "1600 W Merit Pkwy",
      "address_2": "N/A",
      "postal_code": "84095-2416",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.",
      "recall_number": "Z-2296-2025",
      "product_description": "Custom Waste Management Kit:  REF: K10-00353K, K10-01603C, K10-01694Q, K10-05159B, K10-05743, K10-05825AP",
      "product_quantity": "858",
      "reason_for_recall": "High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.",
      "recall_initiation_date": "20250605",
      "center_classification_date": "20250812",
      "report_date": "20250820",
      "code_info": "REF/UDI-DI/Lot(Expiration): K10-00353K/00884450179246/H3127803(1/4/2028), H3166061(2/24/2028), H3172299(3/14/2028); K10-01603C/00884450593639/H3163984(3/2/2028), H3176777(3/23/2028); K10-01694Q/00884450239308/H3116890(10/31/2027), H3152493(2/3/2028), H3177622(3/14/2028); K10-05159B/00884450153598/H3116922(12/23/2027), H3119571(1/4/2028), H3127907(1/11/2028), H3152490(2/10/2028), H3152491(2/10/2028), H3152492(2/10/2028); K10-05743/00884450456767/H3122012(1/6/2028), H3128429(1/13/2028), H3177446(3/16/2028); K10-05825AP/00884450808108/H3112585(11/27/2027)",
      "more_code_info": ""
    }
  ]
}