  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83020",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide-Distribution in the states of CA, FL, GA, IL, KS, MO, PA, RI, TN    Foreign: Canada",
      "recall_number": "Z-2279-2019",
      "product_description": "BD Vacutainer Fluoride Tubes for Blood Alcohol determinations  Part Number: 367001    Product Usage:  Blood Alcohol determinations",
      "product_quantity": "247,000 tubes",
      "reason_for_recall": "Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no longer than two days.  If the sample was stored for more than 2 days, the result for blood alcohol determination might not be accurate (either falsely low or falsely high)",
      "recall_initiation_date": "20190530",
      "center_classification_date": "20190814",
      "termination_date": "20200505",
      "report_date": "20190821",
      "code_info": "Lot Number: 8187663  Exp. Date: 2020/7/31     UDI: (01)30382903670018  (17)200731(10)8187663  (30)0100"
    }
  ]
}