  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80211",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to accounts in CA, IN, KS, ME, and WV, and PR. Foreign distribution to Australia, Bangladesh, Belarus, Belgium, Brazil, Cambodia, Canada, Chile, China, Colombia, Cyprus, Denmark, Dutch Caribbean, Ecuador, Egypt, El Salvador, Ethiopia, Falkland Islands, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Iraq, Ireland, Italy, Jordan, Kenya, Saudi Arabia, Korea, Kuwait, Latvia, Lebanon, Peru, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Morocco, Myanmar, New Zealand, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Oman, Thailand, Netherlands, Turkey, United Arab Emirates, United Kingdom, Ukraine, Vietnam, and Zimbabwe.",
      "recall_number": "Z-2278-2018",
      "product_description": "Randox Liquid Assayed Specific Protein Control - Level 1 (SP Control 1), Cat. no. PS2682",
      "product_quantity": "",
      "reason_for_recall": "The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.",
      "recall_initiation_date": "20180522",
      "center_classification_date": "20180622",
      "termination_date": "20201117",
      "report_date": "20180704",
      "code_info": "All lots.",
      "more_code_info": ""
    }
  ]
}