      "status": "Terminated",
      "city": "Madison",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80173",
      "recalling_firm": "Datex-Ohmeda, Inc.",
      "address_1": "3030 Ohmeda Dr",
      "address_2": "N/A",
      "postal_code": "53718-6704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -  USA (nationwide) Distribution to the states of : AL, AZ, CA, GA, ID, IL, IN, MS, NC, NJ, NK, NV, OH, OK, RI, TX, and WI., and to the countries of :  Albania, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia,  Cyprus, Czech Republic, Ecuador, El Salvador, Estonia, Ethiopia,  Egypt, Finland, France, Germany,  Ghana, Hungary, India, Indonesia, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia,  Mexico, Morocco, Myanmar, Nepal, New Zealand, Nigeria, Norway, Pakistan,  Palestine, Peru, Philippines, Poland, Russia, Saudi Arabia, Senegal, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates,  United Kingdom, Uruguay, and Vietnam",
      "recall_number": "Z-2266-2018",
      "product_description": "CARESCAPE R860 ventilators with software version 10SP05    Product Usage:  The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume. Additional respiratory gas monitoring capabilities are supported through the use of optional GE patient monitoring modules. Not all features are available for all patient types or product configurations. The CARESCAPE R860 ventilator is not a pulmonary function calculation device. The system is designed for facility use, including within facility transport, and should only be used under the orders of a clinician.",
      "product_quantity": "349 (US) + 2,051 (OUS)",
      "reason_for_recall": "Potential for the display processor to experience an  unexpected failure.",
      "recall_initiation_date": "20171208",
      "center_classification_date": "20180620",
      "termination_date": "20210429",
      "report_date": "20180627",
      "code_info": "Lot / Serials #'s: 103841473, 103725811-1, 103725811-2, CBRW01546, CBRW01547, CBRU02110, CBRV00344, CBRV02569, CBRV00298, 103721045, 103808843, CBRV03050, CBRV03054, CBRV03056, CBRV03063,  CBRV03064, CBRV03071, CBRV03072, CBRV03073, CBRW00417,   CBRW00419, CBRW00424, CBRW00441, CBRW00443, CBRW00458,  CBRW00460, CBRW00461, CBRW02306, CBRW02627, CBRW02629,  103721436, CBRU02450, CBRW01735, CBRW01044, CBRW01048,  CBRW01049, CBRW01051, CBRW01052, CBRW01053, CBRW01056,  CBRW01057, CBRW01059, CBRW01061, CBRW01062, CBRW01063,  CBRW01075, CBRW01076, CBRW01077, CBRW01232, CBRW01814,  CBRW01410, CBRW01412, 103798917, CBRW01918, CBRV02578,   CBRW02442, CBRW02449, CBRW01830, CBRW01833, CBRW01838,  CBRW01839, CBRW01849, CBRW01860, CBRW01889, CBRW01895,  CBRW01896, CBRW01897, CBRW01898, CBRW01899, CBRW01900,  CBRW01902, CBRW01903, CBRW01904, CBRW01905, CBRW01906,  CBRW01909, CBRW01932, CBRV01459, 103730032, 103721104,  CBRW01244, CBRW02420, CBRW01347, 103771244, CBRW02661,  CBRU02452, 11495021, 70175736-1, 70175736-10, 70175736-100,  70175736-101, 70175736-102, 70175736-103, 70175736-104, 70175736-105,  70175736-106, 70175736-107, 70175736-108, 70175736-109, 70175736-11,  70175736-110, 70175736-111, 70175736-112, 70175736-113, 70175736-114,  70175736-115, 70175736-116, 70175736-117, 70175736-118, 70175736-119,  70175736-12, 70175736-120, 70175736-121, 70175736-122, 70175736-123,  70175736-124, 70175736-125, 70175736-126, 70175736-127, 70175736-128,   70175736-129, 70175736-13, 70175736-130, 70175736-131, 70175736-132,  70175736-133, 70175736-134, 70175736-135, 70175736-136, 70175736-137,  70175736-138, 70175736-139, 70175736-14, 70175736-140, 70175736-141,  70175736-142, 70175736-143, 70175736-144, 70175736-145, 70175736-146,  70175736-147, 70175736-148, 70175736-149, 70175736-15, 70175736-150,  70175736-151, 70175736-152, 70175736-153, 70175736-154, 70175736-155,  70175736-156, 70175736-157, 70175736-158, 70175736-159, 70175736-16,  70175736-160, 70175736-161, 70175736-162, 70175736-163, 70175736-164,  70175736-165, 70175736-166, 70175736-167, 70175736-168, 70175736-169,  70175736-17, 70175736-170, 70175736-171, 70175736-172, 70175736-173,  70175736-174, 70175736-175, 70175736-176, 70175736-177, 70175736-178,  70175736-179, 70175736-18, 70175736-180, 70175736-181, 70175736-182,  70175736-183, 70175736-184, 70175736-185, 70175736-186, 70175736-187,  70175736-188, 70175736-189, 70175736-19, 70175736-190, 70175736-191,  70175736-192, 70175736-193, 70175736-194, 70175736-195, 70175736-196,  70175736-197, 70175736-198, 70175736-199, 70175736-2, 70175736-20,  70175736-200, 70175736-21, 70175736-22, 70175736-23, 70175736-24,  70175736-25, 70175736-26, 70175736-27, 70175736-28, 70175736-29,  70175736-3, 70175736-30, 70175736-31, 70175736-32, 70175736-33,  70175736-34, 70175736-35, 70175736-36, 70175736-37, 70175736-38,  70175736-39, 70175736-4, 70175736-40, 70175736-41, 70175736-42,  70175736-43, 70175736-44, 70175736-45, 70175736-46, 70175736-47,  70175736-48, 70175736-49, 70175736-5, 70175736-50, 70175736-51,  70175736-52, 70175736-53, 70175736-54, 70175736-55, 70175736-56,  70175736-57, 70175736-58, 70175736-59, 70175736-6, 70175736-60,  70175736-61, 70175736-62, 70175736-63, 70175736-64, 70175736-65,  70175736-66, 70175736-67, 70175736-68, 70175736-69, 70175736-7,   70175736-70, 70175736-71, 70175736-72, 70175736-73, 70175736-74,  70175736-75, 70175736-76, 70175736-77, 70175736-78, 70175736-79,  70175736-8, 70175736-80, 70175736-81, 70175736-82, 70175736-83,   70175736-84, 70175736-85, 70175736-86, 70175736-87, 70175736-88,   70175736-89, 70175736-9, 70175736-90, 70175736-91, 70175736-92,  70175736-93, 70175736-94, 70175736-95, 70175736-96, 70175736-97,  70175736-98, 70175736-99, 103725706, CBRW02482, 103721074,  CBRW01588, CBRW02287, 103721038, CBRW01151, CBRW01152,   CBRW01155, CBRW01156, CBRW01157, CBRW01158, 103737388,  CBRU02113, CBRU02114, CBRU02124, CBRU02126, CBRU02128,  CBRU02135, CBRU02143, CBRU02454, CBRU02456, CBRV01435,   CBRV01436, CBRV01438, CBRV01449, CBRV01451, CBRV01453,   CBRV02039, CBRV02117, CBRV02156, CBRV02179, CBRV02188,  CBRV02199, CBRV02540, CBRV02542, CBRV02563, CBRV02570,  CBRV02571, CBRV02579, CBRV02583, CBRV02584, CBRV01458,  CBRV02025, CBRW00442, CBRW00459, CBRW01055, CBRW01058,  CBRW01074, 11494687, 103722840, 103771194, 103785995-1,   103785995-2, 103785995-3, 103785995-4, 103785995-5, 11466039-1,  11466039-2, 11466039-3, 11466039-4, 11466039-5, CBRW01437, CBRW01448, 11485242-1, 11485242-2, CBRW01112, CBRW01464,  CBRW01797, CBRW01799, CBRW01800, CBRW01810, CBRW01831,  CBRW01842, CBRW01846, CBRW01866, CBRW01867, CBRW01868,   CBRW01869, CBRW01878, CBRW01879, CBRW01880, CBRW01881,  CBRW01882, CBRW01883, CBRW01809, CBRW01934, CBRW01937,  CBRW01946, CBRW01947, CBRW01948, 6527249, 6528751, 6526140-1  6526140-2, 6528030-1, 6528030-2, 6528030-3, CBRW01401, CBRW01439,  CBRW01964, CBRW01965, CBRW01971, CBRW01972, CBRW01974,   CBRW02707, CBRW02735, CBRW02753, CBRV02796, CBRW01328,   CBRW01771, CBRW01783, CBRW01794, CBRW01795, CBRW01796,  CBRW01798, CBRW01802, CBRW01807, CBRW01767, CBRW01768,  CBRW01770, CBRW01772, CBRW02562, CBRW02573, CBRW02574,  CBRW02575, CBRW02576, CBRW02577, CBRW02578, CBRW02579,  CBRW02580, CBRW02581, CBRW02582, CBRW02583, CBRW02584,  CBRW02585, CBRW02586, CBRW02587, CBRW02588, CBRW02598,  CBRW02599, CBRW02600, CBRW02601, CBRW02602, CBRW02603,  CBRW02614, CBRW02615, CBRW02617, CBRW02631, CBRW02632,  CBRW02634, CBRW02635, CBRW02636, CBRW02637, CBRW02638,  CBRW02639, CBRW02640, CBRW02641, CBRW02642, CBRW01147,  CBRW01149, CBRW01573, CBRW01891, CBRW01892, CBRW01893,   CBRW01894, CBRW02605, CBRW02606, CBRW02608, CBRW02609,  CBRW02610, CBRW02612, CBRW02613, CBRW02616, ,CBRW02421,  CBRW02422, CBRW02424, CBRW02425, CBRW01983, CBRW01992,  CBRW01644, CBRW01645, CBRW01658, CBRW01413, CBRW01414,  CBRW01582, CBRW01586, CBRW02301, CBRW02304, CBRW02305,  CBRW02307, CBRW02308, CBRW02309, CBRW02310, CBRW02311,  CBRW02312, CBRW02314, CBRW02506, CBRW02507, CBRW02563,  CBRW02564, CBRW02659, CBRW02660, CBRW02696, CBRW02697,  CBRW02698, CBRW02757, CBRW02758, CBRW02772, CBRW02773,  CBRW02774, CBRW00934, CBRW00935, CBRW00936, CBRW00937,  CBRW00938, CBRW00939, CBRW00940, CBRW01634, CBRW01669,  CBRW01901, CBRW01907, CBRW01908, CBRW01910, CBRW01911,  CBRW01912, CBRW01913, CBRW01914, CBRW01915, CBRW01916,  CBRW01917, CBRW01919, CBRW01920, CBRW01921, CBRW01922,  CBRW01923, CBRW01924, CBRW01925, CBRW01926, CBRW01927,  CBRW00942, CBRW00943, CBRW00944, CBRW02713, CBRW02763,  CBRW02764, CBRW02786, CBRW02810, CBRW02526, CBRU00173,  10190157, 10190159, 10191029, 10193770, 10194582, CBRU00498,  CBRU00509, CBRU00512, CBRU00517, CBRU00522, CBRU00523,  CBRU00540, CBRU00541, CBRU00550, CBRU00551, CBRU00552,  CBRU00553, CBRU00554, CBRU00555, CBRU00559, CBRU00560,  CBRU00561, CBRU00562, CBRU00563, CBRU00565, CBRU00566,  CBRU00567, CBRU00568, CBRU00569, CBRU00570, CBRU00571,  CBRU00572, CBRU00573, CBRU00574, CBRU00575, CBRU00577,  CBRU00579, CBRU00580, CBRU00581, CBRU00593, CBRU00594,  CBRU00595, CBRU00596, CBRU00597, CBRU00598, CBRU00599,  CBRU00600, CBRU00601, CBRU00602, CBRU00603, CBRU00604,  CBRU00605, CBRU00606, CBRU00607, CBRU00608, CBRU00613,  CBRU00614, CBRU00619, CBRU00620, CBRU00621, CBRU00622,  CBRU00624, CBRW00918, CBRW02249, CBRW00919, CBRW01284,  CBRW01570, CBRW01630, 11484134, CBRW01604, CBRW01856,   CBRW01857, CBRW01858, CBRW01375, CBRW01447, CBRW01619,  CBRW01715, CBRW01742, CBRW01746, CBRW01753, CBRW01754,   CBRW01755, CBRW01757, CBRW01758, CBRW01804, CBRW01805,  CBRW01808, CBRW01820, CBRW01821, CBRW01823, CBRW01824,  CBRW01825, CBRW01827, CBRW01834, CBRW01835, CBRW01836,  CBRW02023, CBRW02046, CBRW02048, CBRW02049, CBRW02053,  CBRW02055, CBRW02056, CBRW02060, CBRW02061, CBRW02063,  CBRW02064, CBRW02065, CBRW02066, CBRW02067, CBRW02069,  CBRW02070, CBRW02071, CBRW02072, CBRW02073, CBRW02074,  CBRW02075, CBRW02076, CBRW02077, CBRW02078, CBRW02079,  CBRW02080, CBRW02081, CBRW02082, CBRW02083, CBRW02084,  CBRW02085, CBRW02086, CBRW02087, CBRW02088, CBRW02089,  CBRW02090, CBRW02091, CBRW02092, CBRW02093, CBRW02094,  CBRW02095, CBRW02096, CBRW02097, CBRW02098, CBRW02099,  CBRW02100, CBRW02102, CBRW02105, CBRW02107, CBRW02108,  CBRW02109, CBRW02110, CBRW02114, CBRW02115, CBRW02120,  CBRW02122, CBRW02123, CBRW02124, CBRW02125, CBRW02126,  CBRW02127, CBRW02128, CBRW02129, CBRW02130, CBRW02131,  CBRW02144, CBRW02145, CBRW02146, CBRW02147, CBRW02216,  CBRW02611, CBRW02633, CBRW02648, CBRW02652, CBRW02653,  CBRW02654, CBRW02655, CBRW02656, CBRW02657, CBRW02658,  CBRW02676, CBRW02677, CBRW02678, CBRW02679, CBRW02684,  CBRW02686, CBRW02687, CBRW02688, CBRW02689, CBRW02690,  CBRW02693, CBRW02694, CBRW02695, CBRW02699, CBRW02700,  CBRW02702, CBRW02037, CBRW02051, CBRW02285, CBRW02298,  CBRW02103,CBRW02104, CBRW02143, CBRW01578, CBRW01928,  CBRW01929, CBRW01930, CBRW01933, CBRW01952, CBRW01954,  CBRW02021, CBRW02044, CBRW02045, CBRW02047, CBRW02050,  CBRW02052, CBRW02038, CBRW01991, CBRW02018, CBRW02028,  CBRW02034, CBRW02035, CBRW02036, CBRW00982, CBRW00983,  CBRW00984, CBRW00985, CBRW00986, CBRW01829, CBRW01850,  CBRW01851, CBRW01855, CBRW01859, CBRW01568, CBRW01574,  CBRW01108, CBRW02592, CBRW02593, CBRW02594, CBRW01078,  CBRW01079, CBRW01080, CBRW01081, CBRW01082, CBRW01083,  CBRW01084, CBRW01085, CBRW01086, CBRW01087, CBRW01614,  CBRW01615, CBRW01617, CBRW01621, CBRW01628, CBRW01629,  CBRW01637, CBRW01638, CBRW01639, CBRW01650, CBRW01663,   CBRW01202, CBRW01203, CBRW01207, CBRW01211, CBRW01217,  CBRW01218, CBRW01745, CBRW01585, CBRW01593, CBRW01269,  CBRW01270, CBRW01271, CBRW01272, CBRW01273, CBRW01274,  CBRW01275, CBRW01276, CBRW01277, CBRW01278, CBRW01279,  CBRW01280, CBRW01281, CBRW01282, CBRW01688, CBRW01734,  CBRW01748, CBRW01765, CBRW01766, CBRW01773, CBRW01781,  CBRW01782, CBRW01784, CBRW01785, CBRW01786, CBRW01787,  CBRW01789, CBRW01790, CBRW01791, CBRW01792, CBRW01793,  CBRW01938, CBRW01941, CBRW01942, CBRW01943, ,CBRW01944  CBRW01945, CBRW01949, CBRW01956, CBRW01957, CBRW01958,  CBRW02483, CBRW02484, CBRW02485, CBRW02486, CBRW02487,  CBRW02490, CBRW02493, CBRW02494, CBRW02496, CBRW02497,  CBRW02498, CBRW02499, CBRW02500, CBRW02501, CBRW02502,  CBRW02503, CBRW02504, CBRW02511, CBRW02565, CBRW02566,  CBRW02568, CBRW02590, CBRW02591, CBRW01551, CBRW01564,  CBRW01565, CBRW01566, CBRW01567, CBRW01601, CBRV01842,  11489276, CBRW01571, CBRW01572, CBRW01959, CBRW01961,  CBRW01710, CBRW02423, CBRW02433, CBRW02448, CBRW02548,  CBRW02553, CBRW02559, CBRW02589, CBRW02595, CBRW02596,  CBRW02618, CBRW02619, CBRW02622, CBRW02623, CBRW02625,  CBRW02626, CBRW02646, CBRW02647, CBRW02668, CBRW02672,  CBRW01285, CBRW01587, CBRW01589, CBRW01594, CBRW01596,  CBRW01110, CBRW01120, CBRW01121, CBRW01138, CBRV01741,  CBRW01045, CBRW01046, CBRW01047, CBRW01119, CBRW01136,  CBRW01141, CBRW01713, CBRW01720, CBRW01732, CBRW01743,   CBRW01744, CBRW01756, CBRW01761, CBRW01832, CBRW01862,  CBRW01885, CBRW01886, CBRW02170, CBRW02329, CBRW02330,  CBRW02331, CBRW02332, CBRW02333, CBRW02334, CBRW02335,  CBRW02336, CBRW02337, CBRW02338,  CBRW02339, CBRW02340,  CBRW02341, CBRW02342, CBRW02343, CBRW02344, CBRW02345,  CBRW02346, CBRW02347, CBRW02348, CBRW02351, CBRW02352,  CBRW02353, CBRW02354, CBRW02356, CBRW02357, CBRW02358,  CBRW02374, CBRW02378, CBRW02379, CBRW02380, CBRW02381,  CBRW02382, CBRW02383, CBRW02384, CBRW02385, CBRW02387,  CBRW02388, CBRW02390, CBRW02391, CBRW02392, CBRW02393,  CBRW02394, CBRW02471, CBRW02472, CBRW02473, CBRW02474,  CBRW02475, CBRW02489, CBRW02491, CBRW02492, CBRW02508,  CBRW02510, CBRW02515, CBRW02516, CBRW02525, CBRW02528,  CBRW02533, CBRW02543, CBRW02546, CBRW02572, CBRW02597,  CBRW02620, CBRW02621, CBRW02711, CBRW02746, CBRW02747,  CBRW02752, CBRW02754, CBRW02755, CBRW02803, CBRW02828,  CBRW02844, CBRW02850, CBRW01840, CBRW01884, CBRW01580,  CBRW01848, CBRW01865, CBRW02544, 5447395, 5449444, 5449882,  5450368, 5450806, 5453175, 5456247, 5458189, 5445615-1, 5445615-2,  5445657-1, 5445657-2, 5445657-3, 5445657-4, 5445705-1, 5445705-2,  5445723-1, 5445723-2, 5445815-1, 5445815-2, 5445815-3, 5445816-1,  5445816-2, 5447746-1, 5447746-2, 5447746-3, 5447746-4, 5447746-5,  5448449-1, 5448449-2, 5454469-1, 5454469-2, 5457167-1, 5457167-10,  5457167-11, 5457167-12, 5457167-13, 5457167-14, 5457167-15, 5457167-16,  5457167-17, 5457167-18, 5457167-19, 5457167-2, 5457167-3, 5457167-4,  5457167-5, 5457167-6, 5457167-7, 5457167-8, 5457167-9, 5457943-1,   5457943-2, 5457943-3, 5457943-4, CBRW01092, CBRW01129, CBRW01131,  CBRW01132, CBRW01133, CBRT00331, CBRW01236, CBRW01981,   CBRW02181, CBRW02182, CBRW02184, CBRW02187, CBRW02197,  CBRW02198, CBRW02201, CBRW02203, CBRW02206, CBRW02208,  5447306, 5457519, 11017048, 11017065, 11017073, 11017091, 11017097,  11017109, 11017125, 11017186, 11017371, 11017406, 11017419, 11017424,  11017509, 11017510, 11017514, 11017670, 11017725, 11017795, 11017949,  11019263, 11017185-1, 11017185-2, 11017185-3, 11017185-4, 11017185-5,  11017488-1, 11017488-2, 11017488-3, 11017488-4,  11017488-5, BRW00964,  CBRW00965, CBRW00966, CBRW00967, CBRW01216, CBRW01224,   CBRW01283, CBRW01344, CBRW01345, CBRW01346, CBRW01352,  CBRW01353, CBRW01680, CBRW01861, CBRW01863, CBRW01876,  CBRW01877, CBRW01887, CBRW01888, CBRW01890, CBRW02010,  CBRW02011, CBRW02013, CBRW02101, CBRW02303, CBRW02313,  CBRW02316, CBRW02370, CBRW02371, CBRW02373, CBRW02375,  CBRW02389, CBRW02395, CBRW02396, CBRW02397, CBRW02398,  CBRW02399, CBRW02400, CBRW02401, CBRW02402, CBRW02407,  CBRW02408, CBRW02409, CBRW02410, CBRW02411, CBRW02412,  CBRW02413, CBRW02414, CBRW02415, CBRW02416, CBRW02417,  CBRW02418, CBRW02419, CBRW02426, CBRW02427, CBRW02428,  CBRW02430, CBRW02450, CBRW02538, CBRW02542, CBRW02771,  CBRW02776, CBRW02794, CBRW02795, CBRW02796, CBRW02797,  CBRW02798, CBRW02820, CBRW01548, CBRW01556, CBRW01558,  CBRW01137, CBRW01139, CBRW01140, CBRW01145, CBRW01146,   CBRW01122,  CBRW01123,  CBRW01124,  CBRW01125,  CBRW01126,  CBRW01127,  CBRW01128,  CBRW01130,  CBRU02111,  CBRW01659,  CBRW01391,  CBRW01392,  CBRW01517,  CBRW01518,  CBRW01519,  CBRW01522,  CBRW01523,  CBRW01525,  CBRW01527,  CBRW01528,  CBRW01529,  CBRW01530,  CBRU00192,  CBRU03363,  CBRW01837,  CBRU02907,  CBRW01198,  CBRW01181,  CBRW01182,  CBRW01183,  CBRU02631,  CBRW01332,  CBRW01374,  CBRW01376,  CBRW01377,  CBRW01378,  CBRW01386,  CBRW01387,  CBRW01388,  CBRW01389,  CBRW01390,  CBRW01247,  CBRW01427,  CBRW01428,  CBRV01710,  CBRW01488,  CBRW01489,  CBRW01490,  CBRW02728,  CBRW02759,  6526298,  6526558,  6527101,  6529043,  CBRW01205,  CBRW01206,  CBRW01215,  CBRW01222,  CBRW01223,  CBRW01434,  CBRW01435,  CBRW01463,  CBRW01467,  CBRW01477,  CBRW01503,  CBRW01504,  CBRW01616,  CBRW01626,  CBRW01670,  CBRW01681,  CBRW01683,  CBRW01685,  CBRW01691,  CBRW01711,  CBRW01714,  CBRW01716,  CBRW01717,  CBRW01718,  CBRW01719,  CBRW01721,  CBRW01806,  CBRW01815,  CBRW01816,  CBRW01817,  CBRW01818,  CBRW01819,  CBRW01822,  CBRW01828,  CBRW01843,  CBRW01844,  CBRW01845,  CBRW01852,  CBRW01853,  CBRW01864,  CBRW01870,  CBRW01871,  CBRW01872,  CBRW01873,  CBRW01874,  CBRW01875,  CBRW01935,  CBRW01936,  CBRW01939,  CBRW01940,  CBRW01950,  CBRW01960,  CBRW01966,  CBRW01967,  CBRW01968,  CBRW01970,  CBRW02137,  CBRW02153,  CBRW02154,  CBRW02155,  CBRW02156,  CBRW02159,  CBRW02163,  CBRW02164,  CBRW02167,  CBRW02168,  CBRW02169,  CBRW02171,  CBRW02172,  CBRW02173,  CBRW02177,  CBRW02178,  CBRW02188,  CBRW02317,  CBRW02320,  CBRW02321,  CBRW02322,  CBRW02323,  CBRW02324,  CBRW02325,  CBRW02326,  CBRW02327,  CBRW02328,  CBRW00941,  CBRW01444,  CBRW01442,  CBRW01443,  CBRW01001,  CBRW01002,  CBRW01003,  CBRW01004,  CBRW01006,  CBRW01007,  CBRW01226,  CBRW01228,  CBRW01241,  CBRW01230,  CBRW01231,  CBRW01234,  CBRW01235,  CBRW01237,  CBRW01238,  CBRW00981,  CBRW00988,  CBRW00989,  CBRW00990,  CBRW00998,  CBRW00999,  CBRW01117,  CBRW01118,  CBRW01415,  CBRW01416,  CBRW01417,  CBRW01418,  CBRW01060,  CBRW01433,  CBRW01436,  CBRW01438,  CBRW01440,  CBRW01441,  CBRW01395,  CBRW01400,  CBRW01398,  CBRW01399,  CBRW01173,  CBRW01175,  CBRW01176,  CBRW01177,  CBRW01178,  CBRW01179,  CBRW01419,  CBRW01420,  CBRW01421,  CBRW01422,  CBRW01423,  CBRW01424,  CBRW01425,  CBRW01426,  CBRW01199,  CBRW01201,  CBRW01379,  CBRW01380,  CBRW01381,  CBRW01383,  CBRW01384,  CBRW01396,  CBRW01200,  CBRW01204,  CBRW01618,  CBRW01712,  CBRW01736,  CBRW01751,  CBRW01759,  CBRW01763,  CBRW01764,  CBRW01043,  CBRW01050,  CBRW01054,  CBRW01106,  CBRW01195,  CBRW01196,  CBRW01197,  CBRW01208,  CBRW01209,  CBRW01213,  CBRW01214,  CBRW01219,  CBRW01220,  CBRW01225,  CBRW01227,  CBRW01229,  CBRW01233,  CBRW01239,  CBRW01240,  CBRW01847,  CBRW01969,  CBRW02003,  CBRW02004,  CBRW02006,  CBRW02009,  CBRW02012,  CBRW02015,  CBRW02016,  CBRW02022,  CBRW02027,  CBRW02030,  CBRW02031,  CBRW02054,  CBRW02057,  CBRW02058,  CBRW02059,  CBRW02062,  CBRW02111,  CBRW02113,  CBRW02118,  CBRW00909,  CBRW00910,  CBRW00911,  CBRW00912,  CBRW00913,  CBRW00915,  CBRW00916,,  CBRW00917,  CBRW00920,  CBRW00921,  CBRW00922,  CBRW00923,  CBRW00924,  CBRW00925,,  CBRW00926,  CBRW00927,  CBRW00928,  CBRW00929,  CBRW00930,  CBRW00931,  CBRW00932,  CBRW00933,  CBRW00945,  CBRW00946,  CBRW00947,  CBRW00948,  CBRW00949,  CBRW00950,  CBRW00951,  CBRW00952,  CBRW00953,  CBRW00954,  CBRW00955,  CBRW00956,  CBRW00957,  CBRW00958,  CBRW00959,  CBRW00960,  CBRW01364,  CBRW01365,  CBRW01366,  CBRW01367,  CBRW01931,  CBRW01445,  CBRW01526,  CBRW01549,  CBRW01552,  CBRW01598,  CBRW01613,  CBRW01620,  CBRW01623,  CBRW01627,  CBRW01660,  CBRW01667,  CBRW01668,  CBRW01245,  CBRW01246,  CBRW01248,  CBRW01249,  CBRW01250,  CBRW01251,  CBRW01252,  CBRW01253,  CBRW01254,  CBRW01255,  CBRW01256,  CBRW01257,  CBRW01258,  CBRW01259,  CBRW01260,  CBRW01261,  CBRW01262,  CBRW02302,  CBRW01987,  CBRW01988,  CBRW02005,  CBRW02019,  CBRW01243,  CBRW01148,  CBRW01150,  CBRW01164,  CBRW01185,  CBRW01186,  CBRW01187,  CBRW01188,  CBRW01189,  CBRW01190,  CBRW01191,  CBRW01192,  CBRW01193,  CBRW01194,  CBRW01550,  CBRW01569,  CBRW01576,  CBRW01579,  CBRW01597,  CBRW01611,  CBRW01612,,  CBRW01641,  CBRW01643,  CBRW00997,  CBRW02434,  CBRW02435,  CBRW02436,  CBRW02437,  CBRW02438,  CBRW02439,  CBRW02440,  CBRW02441,  CBRW02444,  6528877,  CBRW01406,  CBRW01407,  CBRW01408,  CBRW01409,  CBRW01411,  CBRW02232,  CBRW02236,  CBRW02238,  CBRW00971,  CBRW00972,  CBRW00973,  CBRW00974,  CBRW00975,  CBRW00976,  CBRW00977,  CBRW00978,  CBRW00979,  CBRW00980,  CBRW01354,  CBRW01473,  CBRW01474,  CBRW01476,  CBRW01042,  6526204-1,  6526204-2,  6526204-3,  CBRW0174,  CBRW01184,  CBRT00611  CBRW01554,  CBRW01562,  CBRW01575,  CBRW01591,  CBRW01605,  CBRW01606,  CBRW01607,,  CBRW01635,  CBRW01553  CBRW01975  CBRW01976  CBRW02527  CBRW01368  CBRW02199  CBRW02200  CBRV02046  CBRW02121  CBRW01098  CBRW01100  CBRW01101  CBRW01103  CBRW01286  CBRW01287  CBRW01487  CBRW01515  CBRW01524  CBRW02191  CBRW02207  CBRW02545  CBRW02558  CBRW01640  CBRW01653  CBRW01654  CBRW01656  CBRW01657  CBRW01664  CBRW01665  CBRW01666  CBRW02429  CBRW01242  CBRW02007  CBRW02008  CBRW01531  CBRW01532  CBRW01533  CBRW01534  CBRW01535  CBRW01536  CBRW01180  CBRW01446  CBRW01449  CBRW01450  CBRW01451  CBRW01452  CBRW01453  CBRW01454  CBRW01455  CBRW01456  CBRW01457  CBRW01458  CBRW01459  CBRW01460  CBRW01461  CBRW01462  CBRW01465  CBRW01466  CBRW01468  CBRW01469  CBRW01470  CBRW01471  CBRW01472  CBRW01475  CBRW01481  CBRW01483  CBRW01484  CBRW01973  CBRW01351  CBRW01491  CBRW01492  CBRW01493  CBRW01494  CBRW01495  CBRW01496  CBRW01497  CBRW01498  CBRW01499  CBRW01500  CBRW01501  CBRW01502  CBRW01513  CBRW01514  CBRW02202  CBRW01595  CBRW01672  CBRW01682  CBRW01686  CBRW01689  CBRW01692  CBRW01694  CBRW01695  CBRW01696  CBRW01697  CBRW01699  CBRW01708  CBRW00962  CBRW00963  CBRW00968  CBRW00969  CBRW00970  CBRW01625  CBRW01633  CBRW01636  CBRW01662  CBRW01671  CBRW01679  CBRW01698  CBRW01700  CBRW01701  CBRW01702  CBRW01703  CBRW01704  CBRW01705  CBRW01706  CBRW01707  CBRW01263  CBRW01264  CBRW01265  CBRW01266  CBRW01267  CBRW01268  CBRU02180  CBRV02754  CBRW00641  5454414  5454413-1  5454413-2  6527173-1  6527173-2  6527173-3  6527173-4  6527173-5  6527173-6  6528692-1  6528692-2  CBRV01531  CBRV01534  CBRV01535  CBRW01951  CBRW01977  CBRW01978  CBRW01985  CBRW01989  CBRW01995  CBRW01996  CBRW01997  CBRW01998  CBRW01999  CBRW02276  CBRW02277  CBRW02278  CBRW02279  CBRW02294  CBRW02295  CBRW02297  CBRW02522  CBRW02532  CBRW02664  CBRW02670  CBRW02673  CBRW02682  CBRW02683  CBRW02685  CBRW02709  CBRW02710  CBRW02715  CBRW02721  CBRW02722  CBRW02723  CBRW02729  CBRW02741  CBRW02742  CBRW02745  CBRW02751  CBRW02756  CBRW02760  CBRW02761  CBRW02767  CBRW02770  CBRW02789  CBRW02811  CBRW02827  CBRW02842  CBRW01540  CBRW01545  CBRW01642  CBRW01622  CBRW01624  CBRW01693  CBRW01728  CBRW01729  CBRW01730  CBRW01775  CBRW01776  CBRW01777  CBRW01778  CBRW01779  CBRW01780  CBRW01984  CBRW01994  CBRW02000  CBRW02001  CBRW02002  CBRW02296  CBRW01762  CBRW02189  CBRW00961  6526481-1  6526481-2  CBRW01813  CBRW02106  CBRW02119  CBRW02607  CBRW02680  CBRW02681  CBRW02724  CBRW02727  CBRW02782  CBRW02847  CBRW02848  CBRW02869  CBRW01429  CBRW01430  CBRW01431  CBRW01432  CBRW02020  CBRW02024  CBRW02029  CBRW02032  CBRW02033  CBRW01337  CBRW02116  CBRW02117  CBRW01980  CBRW02014  CBRW02017  CBRW02025  CBRW02026  CBRW02039  CBRW02040  CBRW01955  CBRW01962  CBRW01963  CBRW01979  CBRW01537  CBRW01538  CBRW01539  CBRW01541  CBRW01542  CBRW01543  CBRW02315  CBRW02319  6529674  CBRW00987  CBRW00992  CBRW00993  CBRW00994  CBRW00995  CBRW00996  CBRW01000  CBRW01005  CBRW01008  CBRW01011  CBRW01012  CBRW01013  CBRW01014  CBRW01015  CBRW01017  CBRW01025  CBRW01397  CBRW01402  CBRW01403  CBRW01404  CBRW01405  CBRW01982  CBRW01986  CBRW01990  CBRW01993  CBRW02160  CBRW02174  CBRW02175  CBRW02176  CBRW02243  CBRW01212  CBRW01221  CBRW01649  CBRW01655  CBRW01675  CBRW01676  CBRW01677  CBRW01678  CBRW01651  CBRW02258  CBRW02259  CBRW02260  CBRW02261  CBRW02262  CBRW02263  CBRW02265  CBRW02266  CBRW01557  CBRW01563  CBRW01577  CBRW01583  CBRW01590  CBRW01600  CBRW01602  CBRW01603  CBRW01610  CBRW01632  CBRW01646  CBRW01647  CBRW01648  CBRW01652  CBRW01673  CBRW01674  CBRW01363  CBRW01369  CBRW01370  CBRW01371  CBRW01372  CBRW01373  CBRW01544  CBRW01555  CBRW01599  CBRW01608  CBRW01609  CBRW02431  CBRW02432  CBRW02443  CBRW02445  CBRW02447  CBRW02547  CBRW02549  CBRW02550  CBRW02551  CBRW02552  CBRW02556  CBRW02557  CBRW02567  CBRW02570  CBRW02571  CBRW01095  CBRW01099  CBRW01288  CBRW01289  CBRW01290  CBRW01291  CBRW01296  CBRW01788  CBRW01803  CBRW01826  CBRW02554  CBRW02555  CBRW02560  CBRW02561  CBRW02644  CBRW02662  CBRW02691  CBRW02692  CBRW02708  CBRW02730  CBRW02731  CBRW02732  CBRW02733  CBRW02734  CBRW02736  CBRW02737  CBRW02738  CBRW02739  CBRW01159  CBRW01760  CBRW01811  CBRW01812  CBRW02604  CBRW01210  CBRW02299  CBRW02300  CBRW01631  CBRW01661  CBRW01687  CBRW01690  CBRW01709  CBRW01722  CBRW01723  CBRW01724  CBRW01725  CBRW01726  CBRW01731  CBRW01747  CBRW01749  CBRW01750  CBRW01752  CBRW01801  CBRW01841  CBRW01854  CBRW02041  CBRW02042  CBRW02043  CBRW00991  CBRW01009  CBRW01010  CBRW01016  CBRW01018  CBRW01026  CBRW01027  CBRW01028  CBRW01029  CBRW01030  CBRW01037  CBRW01038  CBRW01039  CBRW01040  CBRW01041  CBRW01088  CBRW01089  CBRW01090  CBRW01091  CBRW01093  CBRW01094  CBRW01292  CBRW01293  CBRW01294  CBRW01295  CBRW01297  CBRW01298  CBRW01299  CBRW01300  CBRW01301  CBRW01302  CBRW01303  CBRW01304  CBRW01305  CBRW01306  CBRW01307  CBRW01308  CBRW01309  CBRW01310  CBRW01311  CBRW01312  CBRW01313  CBRW01314  CBRW01315  CBRW01316  CBRW01317  CBRW01318  CBRW01319  CBRW01320  CBRW01321  CBRW01322  CBRW01323  CBRW01324  CBRW01325  CBRW01326  CBRW01327  CBRW01329  CBRW01330  CBRW01331  CBRW01333  CBRW01338  CBRW01727  CBRW01733  CBRW01738  CBRW01740  CBRW01769  CBRW01774  CBRW02068  CBRW02112  CBRW02349  CBRW02350  CBRW02355  CBRW02359  CBRW02360  CBRW02361  CBRW02362  CBRW02363  CBRW02364  CBRW02365  CBRW02366  CBRW02367  CBRW02368  CBRW02369  CBRW02372  CBRW02376  CBRW02377  CBRW02386  CBRW02403  CBRW02404  CBRW02446  CBRW02451  CBRW02452  CBRW02453  CBRW02454  CBRW02455  CBRW02456  CBRW02457  CBRW02458  CBRW02459  CBRW02460  CBRW02461  CBRW02462  CBRW02463  CBRW02464  CBRW02465  CBRW02466  CBRW02467  CBRW02468  CBRW02480  CBRW02481  CBRW02512  CBRW02513  CBRW02624  CBRW02628  CBRW02630  CBRW02643  CBRW02645  CBRW02649  CBRW02650  CBRW02651  CBRW02669  CBRW02671  CBRW02674  CBRW02675  CBRW02701  CBRW02703  CBRW02704  CBRW02705  CBRW02706  CBRW02712  CBRW02717  CBRW02718  CBRW02719  CBRW02720  CBRW02725  CBRW02726  CBRW02740  CBRW02743  CBRW02744  CBRW02765  CBRW02766  CBRW02768  CBRW02769  CBRW02775  CBRW02777  CBRW02778  CBRW02779  CBRW02780  CBRW02781  CBRW02783  CBRW02784  CBRW02785  CBRW02790  CBRW02791  CBRW02792  CBRW02793  CBRW02800  CBRW02801  CBRW02802  CBRW02804  CBRW02806  CBRW02809  CBRW02812  CBRW02813  CBRW02814  CBRW02815  CBRW02816  CBRW02817  CBRW02818  CBRW02819  CBRW02821  CBRW02822  CBRW02823  CBRW02824,  CBRW02830,  CBRW02831,  CBRW02832,  CBRW02833,  CBRW02835,  CBRW02836,  CBRW02837,  CBRW02838,  CBRW02839,  CBRW02843,  CBRW02849,  CBRW02851,  CBRW02852,  CBRW02853,  CBRW02854,  CBRW02855,  CBRW02856,  CBRW02857,  CBRW02858,  CBRW02859,  CBRW02860,  CBRW02861,  CBRW02862,  CBRW02868,  CBRW02870,  CBRW02871,  CBRW02872,  CBRW02874,  CBRW02878,  CBRW02880,  CBRW02881,  CBRW02882,  CBRW02883,  CBRW02884,  CBRW02885,  CBRW02886,  CBRW02887,  CBRW02888,  CBRW02889,  CBRW02890,  CBRW02891,  CBRW02892,  CBRW02893,  CBRW02895,  CBRW02896,  CBRW02902,  CBRW02903,  CBRW02907,  CBRW02909,  CBRW03005,  CBRW03006,  CBRW03007,  CBRW03013,  CBRW03014,  CBRW01478,  CBRW01479,  CBRW01684,  CBRW01737,  CBRW01739,  CBRW01142,  CBRW01143,  CBRW01144,  CBRW01334,  CBRW01335,  CBRW01336,  CBRW01339,  CBRW01340,  CBRW01341,  CBRW01342,  CBRW01343,  CBRW01348,  CBRW01560,  CBRW01561,  CBRW01581,  CBRW01584,  CBRW01592,  CBRW01953,  CBRW02318,  CBRW01096,  CBRW01097,  CBRW01102,  CBRW01104,  CBRW01105,  CBRW01107,  CBRW01109,  CBRW01111,  CBRW01113,  CBRW01114,  CBRW01115,  CBRW01116,  CBRW01134,  CBRW01135,  CBRW01153,  CBRW01154,  CBRW01160,  CBRW01161,  CBRW01162,  CBRW01163,  CBRW01165,  CBRW01166,  CBRW01167,  CBRW01168,  CBRW01169,  CBRW01170,  CBRW01171,  CBRW01172,  CBRW01174,  CBRW01349,  CBRW01350,  CBRW01355,  CBRW01356,  CBRW01357,  CBRW01358,  CBRW01359,  CBRW01360,  CBRW01361,  CBRW01362,  CBRW01382,  CBRW01385,  CBRW01393,  CBRW01394,  CBRW01480,  CBRW01482,  CBRW01485,  CBRW01486,  CBRW01505,  CBRW01506,  CBRW01507,  CBRW01508,  CBRW01509,  CBRW01510,  CBRW01511,  CBRW01512,  CBRW01516,  CBRW01520,  CBRW01521,  CBRW02132,  CBRW02133,  CBRW02134,  CBRW02135,  CBRW02136,  CBRW02138,  CBRW02139,  CBRW02140,  CBRW02141,  CBRW02142,  CBRW02148,  CBRW02149,  CBRW02150,  CBRW02151,  CBRW02152,  CBRW02157,  CBRW02158,  CBRW02161,  CBRW02162,  CBRW02165,  CBRW02166,  CBRW02179,  CBRW02180,  CBRW02183,  CBRW02185,  CBRW02186,  CBRW02190,  CBRW02192,  CBRW02193,  CBRW02194,  CBRW02195,  CBRW02196,  CBRW02204,  CBRW02205,  CBRW02209,  CBRW02210,  CBRW02211,  CBRW02212,  CBRW02213,  CBRW02214,  CBRW02215,  CBRW02217,  CBRW02218,  CBRW02219,  CBRW02220,  CBRW02221,  CBRW02222,  CBRW02223,  CBRW02224,  CBRW02225,  CBRW02226,  CBRW02227,  CBRW02228,  CBRW02229,  CBRW02230  CBRW02231,  CBRW02233,  CBRW02234,  CBRW02235,  CBRW02237,  CBRW02239,  CBRW02240,  CBRW02241,  CBRW02242,  CBRW02244,  CBRW02245,  CBRW02246,  CBRW02247,  CBRW02248,  CBRW02250,  CBRW02251,  CBRW02252,  CBRW02253,  CBRW02254,  CBRW02255,  CBRW02256,  CBRW02257,  CBRW02264,  CBRW02267,  CBRW02268,  CBRW02269,  CBRW02270,  CBRW02271,  CBRW02272,  CBRW02273,  CBRW02274,  CBRW02275,  CBRW02280,  CBRW02281,  CBRW02282,  CBRW02283,  CBRW02284,  CBRW02286,  CBRW02288,  CBRW02289,  CBRW02290,  CBRW02291,  CBRW02292,  CBRW02293,  CBRW02405,  CBRW02406,  CBRW02569,  CBRW02663,  CBRW02665,  CBRW02666,  CBRW02667,"
    }
  ]
}