{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92202",
      "recalling_firm": "Philips North America",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2296",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam",
      "recall_number": "Z-2261-2023",
      "product_description": "Incisive CT Plus-Computed Tomography X-Ray System   Model: 728149",
      "product_quantity": "6 units OUS",
      "reason_for_recall": "Metal mounting box on the rotating scanner on rotor  (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down",
      "recall_initiation_date": "20230425",
      "center_classification_date": "20230728",
      "report_date": "20230809",
      "code_info": "UDI-DI: (01)00884838103474(21)  Serial Numbers: 530408 530437 530474 530320 530332 530430"
    }
  ]
}