  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92681",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "AR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan",
      "recall_number": "Z-2254-2023",
      "product_description": "Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions.  Item Number: 00235710106",
      "product_quantity": "75 units",
      "reason_for_recall": "Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.",
      "recall_initiation_date": "20230627",
      "center_classification_date": "20230728",
      "report_date": "20230809",
      "code_info": "Lot Number/UDI: 64254511 /(01)0089024055940(10)64254511; 64296571/(01)0089024055940(10)64296571; 65379087/(01)0089024055940(10)65379087; 65379093/(01)0089024055940(10)65379093; 65379096/(01)0089024055940(10)65379096"
    }
  ]
}