  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Ludwigsstadt",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94749",
      "recalling_firm": "WOM World Of Medicine AG",
      "address_1": "Alte Poststr. 11",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2243-2024",
      "product_description": "WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc.  REF Numbers: T0449-01",
      "product_quantity": "1700 units",
      "reason_for_recall": "Incorrect transponder data was written on the RFID tags  triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery",
      "recall_initiation_date": "20240524",
      "center_classification_date": "20240701",
      "report_date": "20240710",
      "code_info": "UDI: 04056702003678 Lot Numbers:  4030902, 4030903"
    }
  ]
}