  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Milpitas",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97141",
      "recalling_firm": "Becton, Dickinson and Company, BD Bio Sciences",
      "address_1": "155 N Mccarthy Blvd",
      "address_2": "N/A",
      "postal_code": "95035-5102",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of Tennessee, New York, Massachusetts and the countries of Argentina, Belgium, Botswana, Brazil, Canada, France, Germany, Lithuania, Peru, Philippines, United Kingdom.",
      "recall_number": "Z-2238-2025",
      "product_description": "BD Trucount Controls, 30 Test - REF: 340335",
      "product_quantity": "629 kits",
      "reason_for_recall": "Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.",
      "recall_initiation_date": "20250708",
      "center_classification_date": "20250801",
      "report_date": "20250813",
      "code_info": "Lot: 85708 UDI: 00382903403356 *Distributed outside the US",
      "more_code_info": ""
    }
  ]
}