      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66037",
      "recalling_firm": "Smith & Nephew, Inc. Endoscopy Division",
      "address_1": "150 Minuteman Drive",
      "address_2": "N/A",
      "postal_code": "01810-1031",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide)  Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.",
      "recall_number": "Z-2231-2013",
      "product_description": "TWINFIX Ultra HA 5.5 mm Suture Anchor with 2 ULTRABRAID Suture, Blue / Black; Part Number:  72202626    Biodegradable suture anchor",
      "product_quantity": "3827 US",
      "reason_for_recall": "Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.",
      "recall_initiation_date": "20130806",
      "center_classification_date": "20130917",
      "termination_date": "20160420",
      "report_date": "20130925",
      "code_info": "50324715  50324960  50326833  50336719  50336768  50339643  50339651  50341854  50344396  50348650  50349838  50350841  50351449  50353295  50360151  50363875  50368586  50372045  50374017  50375716  50378770  50381624  50387868  50390937  50395330  50397700  50402973  50404638  50406692  50408058  50412954  50415137  50416799  50418989  50429269  50429640  50429642  50429832  50431134  50439753  50439754  50442740  50443625",
      "more_code_info": ""
    }
  ]
}