  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94239",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, CA, CO, CT, FL, IL, IN, MD, MN, MT, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WY.",
      "recall_number": "Z-2229-2024",
      "product_description": "Nonsterile convenience kits:    a.  Medline Delivery, Pack #DYNJ901130I, containing a Pure Pouch with component Reorder #DYND04000, O.R. Scissors, S/B, 5.5 Sterile.    b.  Medline C-Section, Pack DYNJ902938J, containing a Pure Pouch with component Reorder #DYND04000, O.R. Scissors, S/B, 5.5 Sterile.",
      "product_quantity": "7,673 total non-sterile kits",
      "reason_for_recall": "Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.",
      "recall_initiation_date": "20240314",
      "center_classification_date": "20240628",
      "report_date": "20240710",
      "code_info": "a.  Pack #DYNJ901130I - Lot numbers 23GMH075, exp. 7/31/2024, and 23GMD605, exp. 7/31/2024, UDI-DI each-10193489946055 and UDI-DI case-40193489946056.  b.  Pack DYNJ902938J - Lot numbers 23LMD530, 23IMG375, 23HMB336, 23GMF342, 23EMJ935, and 23EMK254, all exp. date 8/31/2024, UDI-DI each-10193489946116, UDI-DI case 40193489946117."
    }
  ]
}