{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98603",
      "recalling_firm": "Becton, Dickinson and Company",
      "address_1": "17200 Laguna Canyon Rd",
      "address_2": "",
      "postal_code": "92618-5403",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.",
      "recall_number": "Z-2202-2026",
      "product_description": "Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 831F75/831F75P, TS105F5, 774F75, 777F8, AIQSGF8;",
      "product_quantity": "589,223",
      "reason_for_recall": "Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.",
      "recall_initiation_date": "20260408",
      "center_classification_date": "20260515",
      "report_date": "20260527",
      "code_info": "Model/UDI-DI/Lot(Expiration): 131F7P/00690103000412/65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026),65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026),65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026), 66010507(9/13/2026), 66010508(9/14/2026), 66023074(9/30/2026), 66033088(10/5/2026), 66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026), 66120312(10/30/2026), 66120313(10/31/2026), 66127027(11/15/2026), 66127028(11/20/2026), 66132570(11/21/2026), 66132571(11/25/2026), 66132572(11/27/2026), 66140447(11/2/2026), 66140448(11/5/2026), 66140449(11/6/2026), 66140450(11/7/2026), 66142269(11/28/2026), 66142274(11/29/2026), 66151008(12/20/2026), 66160820(11/30/2026), 66160821(12/3/2026), 66160822(12/4/2026), 66160823(12/5/2026), 66174935(12/6/2026), 66186389(12/9/2026), 66186390(12/10/2026), 66186391(12/11/2026), 66186393(12/19/2026), 66214346(12/12/2026), 66214347(12/17/2026), 66214348(12/13/2026), 66214349(12/16/2026), 66225263(1/24/2027), 66230833(1/27/2027), 66230834(1/28/2027), 66231174(1/29/2027), 66236300(2/3/2027), 66243301(1/30/2027), 66243302(2/1/2027), 66254697(2/4/2027), 66254698(2/10/2027), 66273573(2/11/2027), 66273574(2/12/2027), 66283862(2/8/2027), 66296222(3/5/2027), 66315689(3/6/2027), 66351811(3/8/2027), 66351812(3/11/2027), 66351813(3/12/2027), 66378395(3/13/2027), 66378396(3/18/2027), 66384489(3/18/2027), 66414674(3/22/2027), 66438179(4/3/2027), 66438197(4/12/2027), 66438198(5/24/2027), 66438199(5/27/2027), 66438200(4/10/2027), 66439368(3/26/2027), 66439369(3/31/2027), 66439370(4/2/2027), 66439371(4/7/2027), 66439372(4/9/2027), 66488597(5/28/2027), 66521743(4/16/2027), 66529338(4/15/2027), 66529339(4/21/2027), 66529340(4/22/2027), 66529341(4/23/2027), 66529345(5/23/2027), 66544240(4/24/2027), 66594945(5/14/2027), 66597237(4/29/2027), 66597238(4/30/2027), 66597239(5/10/2027), 66601166(5/2/2027), 66601167(5/8/2027), 66601168(5/12/2027), 66601169(5/13/2027), 66601170(5/14/2027), 66601171(5/29/2027), 66642726(5/9/2027), 66692566(6/6/2027), 66704724(6/4/2027), 66710839(6/2/2027), 66710840(6/3/2027), 66739812(6/13/2027), 66739813(6/18/2027), 66739814(6/17/2027), 66739910(6/24/2027), 66739911(7/17/2027), 66739912(7/9/2027), 66739913(6/16/2027), 66739916(6/23/2027), 66740114(7/17/2027), 66758828(7/10/2027), 66767282(6/28/2027), 66769651(7/18/2027), 66782067(7/21/2027), 66826029(7/23/2027), 66826030(7/24/2027), 66832079(7/25/2027), 66844048(7/31/2027), 66873884(8/2/2027), 66873885(8/5/2027), 66873886(8/18/2027), 66895223(8/20/2027), 66938771(8/22/2027), 66938772(8/26/2027), 66938773(8/27/2027), 66964710(9/1/2027), 66964711(9/4/2027), 66982688(9/26/2027), 67025469(10/20/2027), 67047800(10/17/2027), 67047801(10/21/2027), 67068446(10/28/2027), 67092101(10/28/2027), 67098252(10/30/2027), 67098253(11/3/2027), 67199591(11/12/2027), 67199595(11/13/2027), 67199596(11/11/2027), 67199617(11/21/2027), 67199618(11/22/2027), 67217934(1/7/2028), 67217935(12/4/2027), 67266297(1/12/2028), 67275624(12/20/2027), 67275626(12/3/2027), 67322713(1/13/2028), 67322714(1/16/2028), 67357488(1/29/2028), 67395549(2/5/2028), 67395555(1/28/2028), 66739917(6/12/2027);  131F7/00690103000405/ 65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026), 65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026), 65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026) ,66010507(9/13/2026) ,66010508(9/14/2026) ,66023074(9/30/2026) ,66033088(10/5/2026) ,66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026), 66120312(10/30/2026), 66120313(10/31/2026), 66127027(11/15/2026), 66127028(11/20/2026),66132570(11/21/2026), 66132571(11/25/2026), 66132572(11/27/2026), 66140447(11/2/2026), 66140448(11/5/2026), 66140449(11/6/2026),66140450(11/7/2026), 66142269(11/28/2026), 66142274(11/29/2026), 66151008(12/20/2026), 66160820(11/30/2026), 66160821(12/3/2026), 66160822(12/4/2026), 66160823(12/5/2026), 66174935(12/6/2026), 66186389(12/9/2026), 66186390(12/10/2026),66186391(12/11/2026), 66186393(12/19/2026), 66214346(12/12/2026), 66214347(12/17/2026), 66214348(12/13/2026),66214349(12/16/2026), 66225263(1/24/2027), 66230833(1/27/2027), 66230834(1/28/2027), 66231174(1/29/2027), 66236300(2/3/2027), 66243301(1/30/2027), 66243302(2/1/2027), 66254697(2/4/2027), 66254698(2/10/2027), 66273573(2/11/2027),66273574(2/12/2027), 66283862(2/8/2027),66296222(3/5/2027), 66315689(3/6/2027), 66351811(3/8/2027), 66351812(3/11/2027), 66351813(3/12/2027), 66378395(3/13/2027), 66378396(3/18/2027), 66384489(3/18/2027), 66414674(3/22/2027),66438179(4/3/2027), 66438197(4/12/2027), 66438198(5/24/2027),66438199(5/27/2027), 66438200(4/10/2027), 66439368(3/26/2027), 66439369(3/31/2027), 66439370(4/2/2027), 66439371(4/7/2027), 66439372(4/9/2027), 66488597(5/28/2027),66521743(4/16/2027), 66529338(4/15/2027), 66529339(4/21/2027), 66529340(4/22/2027), 66529341(4/23/2027),66529345(5/23/2027),66544240(4/24/2027), 66594945(5/14/2027), 66597237(4/29/2027),66597238(4/30/2027), 66597239(5/10/2027), 66601166(5/2/2027), 66601167(5/8/2027), 66601168(5/12/2027), 66601169(5/13/2027), 66601170(5/14/2027), 66601171(5/29/2027), 66642726(5/9/2027), 66692566(6/6/2027), 66704724(6/4/2027), 66710839(6/2/2027), 66710840(6/3/2027), 66739812(6/13/2027),66739813(6/18/2027), 66739814(6/17/2027), 66739910(6/24/2027), 66739911(7/17/2027), 66739912(7/9/2027), 66739913(6/16/2027), 66739916(6/23/2027), 66740114(7/17/2027), 66758828(7/10/2027), 66767282(6/28/2027), 66769651(7/18/2027), 66782067(7/21/2027), 66826029(7/23/2027),66826030(7/24/2027), 66832079(7/25/2027), 66844048(7/31/2027), 66873884(8/2/2027), 66873885(8/5/2027), 66873886(8/18/2027), 66895223(8/20/2027), 66938771(8/22/2027), 66938772(8/26/2027), 66938773(8/27/2027), 66964710(9/1/2027), 66964711(9/4/2027), 66982688(9/26/2027), 67025469(10/20/2027), 67047800(10/17/2027), 67047801(10/21/2027), 67068446(10/28/2027), 67092101(10/28/2027), 67098252(10/30/2027), 67098253(11/3/2027), 67199591(11/12/2027), 67199595(11/13/2027), 67199596(11/11/2027), 67199617(11/21/2027), 67199618(11/22/2027), 67204649(12/4/2027), 67217934(1/7/2028), 67217935(12/4/2027), 67266297(1/12/2028), 67268374(12/3/2027), 67275624(12/20/2027), 67275626(12/3/2027), 67322713(1/13/2028), 67322714(1/16/2028), 67357488(1/29/2028), 67395549(2/5/2028), 67395555(1/28/2028);  131F7J/67268374(12/3/2027);  141F7/00690103000504/ 65891632(8/5/2026), 65891637(8/20/2026), 65916934(8/19/2026), 65916935(8/26/2026), 65944795(8/27/2026), 65955863(8/27/2026), 65973744(9/9/2026), 65988572(9/18/2026), 65997483(10/1/2026), 66023075(10/4/2026), 66023076(10/28/2026), 66033077(10/25/2026), 66061553(10/26/2026), 66273575(2/13/2027), 66315690(3/8/2027), 66378397(3/20/2027), 66414675(3/28/2027), 66414681(5/30/2027), 66465246(4/23/2027), 66552320(4/29/2027), 66594946(4/28/2027), 66623671(5/1/2027), 66644932(5/16/2027), 66739789(6/26/2027), 66767253(6/30/2027), 66809856(7/4/2027), 66809857(7/7/2027), 66812062(7/8/2027), 66826050(7/30/2027), 66873887(7/29/2027), 66945954(8/26/2027), 67014463(12/19/2026), 67056072(9/22/2027);  151F7/00690103000566/ 66001841(8/29/2026), 66003074(9/14/2026), 66003075(9/7/2026), 66017563(9/21/2026), 66081903(10/3/2026), 66081904(11/1/2026), 66106780(10/25/2026), 66108265(11/7/2026), 66174936(12/17/2026), 66351108(3/6/2027), 66351109(3/12/2027), 66351111(3/21/2027), 66384522(3/29/2027), 66439364(4/4/2027), 66471244(4/11/2027), 66544241(4/24/2027), 66552321(5/2/2027), 66594947(5/9/2027), 66615747(5/14/2027), 66616064(6/27/2027), 66643585(5/21/2027), 66670993(8/7/2027), 66710846(6/6/2027), 66758827(8/1/2027), 66809858(8/6/2027), 66809859(9/11/2027), 66809860(9/20/2027), 66889370(8/26/2027), 66972526(10/3/2027), 67197204(11/20/2027), 67197205(11/28/2027); 834F75/00690103151756/ 66149781(11/20/2026), 66169452(11/21/2026), 66169453(11/22/2026), 66169454(11/22/2026), 66225269(1/17/2027), 66292951(2/6/2027), 66295865(2/25/2027), 66325010(2/21/2027), 66358925(2/25/2027), 66360368(2/24/2027), 66384531(3/21/2027), 66384532(3/21/2027), 66384533(3/31/2027), 66419531(3/15/2027), 66597740(4/25/2027), 66704730(6/6/2027), 66710835(10/16/2027), 66739754(7/2/2027), 66748559(6/11/2027), 66748561(6/10/2027), 66829581(7/5/2027), 66832094(8/1/2027), 66894451(7/31/2027), 66894452(8/1/2027), 66987645(8/30/2027), 67025475(10/13/2027), 67045003(9/17/2027), 67047807(10/14/2027), 67075111(9/27/2027), 67197192(11/11/2027), 67205219(11/18/2027);  096F6, 096F6P/00690103000245, 00690103218602/66687312(6/9/2027), 66687922(6/10/2027), 66687924(6/13/2027), 66780053(6/26/2027), 66822159(9/22/2027);   C144F7/00690103005134/66822158(7/15/2027);   782F75M/00690103197303/66725374(6/19/2027);   132F5/00690103000467/66687936(6/9/2027), 66467271(6/9/2027);   C146F7/00690103153132;  831F75P/831F75-00690103003031, 00690103218282/66175089(11/26/2026), 66175090(12/2/2026), 66175091(12/2/2026), 66175092(12/3/2026), 66175093(11/26/2026), 66175094(11/27/2026), 66175095(11/28/2026), 66175096(11/29/2026), 66175097(11/29/2026), 66175103(12/4/2026), 66175104(12/5/2026), 66175105(12/5/2026), 66175120(12/9/2026), 66175121(12/10/2026), 66175122(12/11/2026), 66208306(12/12/2026), 66208323(12/13/2026), 66211381(12/16/2026), 66211385(1/8/2027), 66211386(1/9/2027), 66225264(1/13/2027), 66225265(1/14/2027), 66225266(1/15/2027), 66225267(1/16/2027), 66225268(1/22/2027), 66268847(1/23/2027), 66278675(2/5/2027), 66283844(2/14/2027), 66283845(2/26/2027), 66283846(2/20/2027), 66296223(3/3/2027), 66298719(3/1/2027), 66301299(2/17/2027), 66301300(2/19/2027), 66439365(4/3/2027), 66439366(4/5/2027), 66439367(4/11/2027), 66488586(3/28/2027), 66529347(5/31/2027), 66554378(4/12/2027), 66591113(5/2/2027), 66660610(5/20/2027), 66664653(5/21/2027), 66667556(5/17/2027), 66667557(5/19/2027), 66704725(6/5/2027), 66704758(6/27/2027), 66739799(7/15/2027), 66758829(7/11/2027), 66759128(6/12/2027), 66767284(6/25/2027), 66782068(7/28/2027), 66804400(6/30/2027), 66804401(7/2/2027), 66804402(7/3/2027), 66826032(7/26/2027), 66873891(8/8/2027), 66934651(8/23/2027), 66934677(8/29/2027), 66964714(8/25/2027), 66972587(9/18/2027), 66972588(9/22/2027), 67003703(9/30/2027), 67003704(10/1/2027), 67005151(9/9/2027), 67005156(9/3/2027), 67005540(10/7/2027), 67005541(10/9/2027), 67005542(9/10/2027), 67005543(9/11/2027), 67005544(9/23/2027), 67005545(10/10/2027), 67184885(11/1/2027);  TS105F5/00690103018295;  774F75/00690103146998/66521746(5/7/2027), 66844440(7/8/2027), 66844441(7/9/2027), 66844442(7/10/2027), 66844443(7/11/2027), 66844444(7/11/2027), 66844445(7/12/2027), 66844446(7/14/2027), 66844447(7/15/2027), 66844448(7/15/2027), 66844449(7/16/2027), 66844450(7/16/2027), 66844451(7/17/2027), 66844452(7/18/2027), 66844453(8/4/2027), 66919298(8/6/2027), 66919299(8/20/2027), 66920022(8/5/2027);  777F8/00690103146554/66417428(4/21/2027), 66417429(4/22/2027), 66417430(4/22/2027), 66419577(4/11/2027), 66419578(4/12/2027), 66419579(4/14/2027), 66419580(4/14/2027), 66419581(4/15/2027), 66419582(4/15/2027), 66419583(4/16/2027), 66419584(4/16/2027), 66419585(4/21/2027), 66438181(4/23/2027), 66465249(4/25/2027), 66523321(4/23/2027), 66523328(4/24/2027), 66523330(4/28/2027), 66523331(4/25/2027), 66544242(4/30/2027), 66568201(5/7/2027), 66590392(5/12/2027), 66600645(4/28/2027), 66600646(4/29/2027), 66600647(4/29/2027), 66600648(4/30/2027), 66600649(4/30/2027), 66615750(5/1/2027), 66615751(5/3/2027), 66615752(5/19/2027), 66615753(5/20/2027), 66619148(5/2/2027), 66643587(5/22/2027), 66643588(5/23/2027), 66643589(5/26/2027), 66643590(5/27/2027), 66643591(5/27/2027), 66643592(5/28/2027), 66643593(5/29/2027), 66643594(5/30/2027), 66643595(6/2/2027), 66643596(6/3/2027), 66646363(5/12/2027), 66646364(5/13/2027), 66646365(5/14/2027), 66646366(5/15/2027), 66646367(5/16/2027), 66646368(5/16/2027), 66660611(6/3/2027), 66660612(6/4/2027), 66660613(6/4/2027), 66660614(6/5/2027), 66660615(6/5/2027), 66660616(6/7/2027), 66687707(6/7/2027), 66687708(6/9/2027), 66687805(6/9/2027), 66694839(7/5/2027), 66704726(6/10/2027), 66704727(6/11/2027), 66704728(6/11/2027), 66704729(6/12/2027), 66764452(6/12/2027), 66764453(6/13/2027), 66764454(6/16/2027), 66764455(6/16/2027), 66764456(6/17/2027), 66764457(6/17/2027), 66769776(6/18/2027), 66769777(6/23/2027), 66782069(7/7/2027), 66800668(6/23/2027), 66804171(6/28/2027), 66928758(8/7/2027), 67142955(10/30/2027), 67142972(11/14/2027), 67143094(11/27/2027);  AIQSGF8/00690103217094/66414809(3/16/2027), 66487805(5/11/2027), 66417072(3/22/2027), 66493141(5/18/2027), 66493138(5/12/2027), 66417073(3/23/2027), 66437580(3/28/2027);",
      "more_code_info": ""
    }
  ]
}